2.3 Power Supply
See
Safety and Installation Guide 124M8686
for full details.
AC Input
DC Input (Option ADTS405 F only)
2.3.1 405R Rack-mounted version
1.
The unit must be connected to the correct electrical power supply as stated, beside the
power connectors on the rear panel.
2.
Install an easily accessible isolator in the power supply circuit.
3.
The power supply connector on the rear panel of the unit will not be accessible when fitted
in the equipment rack.
Note:
The power supply switch on the front panel is not classified as a
disconnecting device.
On the ADTS 405R, two 20mm High Breaking Capacity fuses are located in the holder mounted on
the power input connector on the rear of the rack and protect the entire unit. The fuses are
connected in the live and neutral supply circuit and are: T5A/250V HRC type. Only replace the fuses
with the correct type, see Figure 2-1 item 11 section 4-4.
2.3.2 405F Flightline version
1.
The power switch on the front panel is not a power isolator as defined by EN61010. To
isolate the power supply disconnect the power supply connector.
Fuses AC Power
The fuses fitted in the fuse holders on the front panel are: T5A/250V HRC type. Only replace a fuse
with the correct type, see Figure 2-1, item 11. To replace a fuse see Section 4-4.
Fuses DC Power
The fuses fit in the fuse holders on the front panel are either: T20.0A/250V HRC type (on the left) or
T32A/250V HRC type (on the right). The DC power option (nominal 28 V DC), can be an alternative
power supply using an additional connector located on the pump front panel. Only replace a fuse
with the correct type, see Figure 2-1, item 11 To replace a fuse see, Section 4-4.
AC power supply
100 / 120 / 230V ~ 50 / 60Hz
115V ~ 400Hz 500VA
DC power supply
24 - 30V
500VA
Содержание Druck ADTS 405 R/F Mk2
Страница 2: ......
Страница 10: ...viii ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Страница 14: ...xii ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Страница 16: ...xiv ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Страница 18: ...xvi ADTS 405 Mk2 Instruction Manual English 2015 General Electric Company All rights reserved ...
Страница 121: ......