Chapter 5. Maintenance
52 | ADTS 405 R/F Mk2 Instruction Manual-English
© 2017 General Electric Company – All Rights Reserved.
5.4.3.1 Replacing ADTS 405F Mk2 AC Fuses
The ADTS 405F Mk2 has two 20 mm High Breaking Capacity (HBC) fuses for the AC supply.
They are located in the holder mounted on the pump-rack front panel. They protect the entire
unit. The fuses are connected in the live and neutral supply circuit. For fuse specification, see
Table 8-2 on page 102.
Figure 5-2: ADTS 405F Mk2 AC Fuse Positioning
a.
Set the power switch to OFF.
b.
Isolate the power supply to the instrument and disconnect the power supply
connector (1).
c.
Remove the fuse holders (2) from the front panel and replace the fuse cartridges (3).
d.
Refit the fuse holders (2) in the front panel.
e.
Connect the power supply connector and switch on the power supply.
f.
Set the unit power switch to ON.
Note:
If the fuse blows immediately after switch-on contact the manufacturer or Service
Agent.
WARNING
Only replace the fuses with the correct type.
2
3
1