User Manual
Druck DPI620G-IS
29 July 2016
116M5464 Revision -
Page 219 of 248
Table 10-2 Current (measure) error limits
Applied
mA
Calibrator
uncertainty
(mA)
Permitted
DPI620G-IS
error (mA)
±55
0.003
±0.0055
±25
0.0025
±0.0040
±20
0.00063
±0.0022
±10
0.00036
±0.0016
±5
0.00025
±0.0013
0 (open circuit)
0.0002
±0.0010
Table 10-3 Current (source) error limits
Source mA
Calibrator
uncertainty
(mA)
Permitted
DPI620G-IS
error (mA)
0.2
0.00008
±0.00011
6
0.00023
±0.00132
12
0.00044
±0.00156
18
0.0065
±0.00180
24
0.0012
±0.00204