Thickness profiles with the B-scan
5 Operation
DMS Go+
Edition 4 (05/2014)
5-43
Switching the B-scan display on
You can switch the B-scan display on any time.
The display of the thickness profile starts automatically
at the first successful coupling after switching the B-
scan display on.
Simultaneously with the B-scan display, the DMS Go+
determines the minimum reading detected during the
measurements. At the end of the selected display time,
this value is displayed above the B-scan, and you can
save the reading by pressing the function key 2 (func-
tion
SEND
).
Note
The processes of saving readings and docu-
menting measurement results are described
in detail in chapter 6.
– In the function group
MEASURE
, select the function
MODE
.
– Choose the option
B-SCAN
in order to switch the B-
scan display on.
Setting the display time
The display time (in seconds) defines the period of time
during which the thickness profile should be recorded.
At the end of this time (maximum 30 seconds), the in-
strument stops the measurements.
The scaling of the X-axis in the B-scan corresponds to
the selected display time.
– In the function group
MEASURE
, select the function
TIME
.
– Set the required display time.
Содержание DMS Go+
Страница 1: ...DMS Go Technical Reference and Operating Manual ...
Страница 3: ...DMS Go Edition 4 05 2014 0 3 Function groups and functions ...
Страница 4: ...0 4 Edition 4 05 2014 DMS Go Function groups and functions cont d ...
Страница 8: ...0 8 Edition 4 05 2014 DMS Go ...
Страница 17: ...DMS Go Edition 4 05 2014 1 1 Introduction 1 ...
Страница 41: ...DMS Go Edition 4 05 2014 2 1 Standard package and accessories 2 ...
Страница 54: ...2 Standard package and accessories Recommended accessories 2 14 Edition 4 05 2014 DMS Go ...
Страница 55: ...DMS Go Edition 4 05 2014 3 1 Initial start up 3 ...
Страница 67: ...DMS Go Edition 4 05 2014 4 1 Principles of operation 4 ...
Страница 78: ...4 12 Edition 4 05 2014 DMS Go ...
Страница 79: ...DMS Go Edition 4 05 2014 5 1 Operation 5 ...
Страница 133: ...DMS Go Edition 4 05 2014 6 1 Data Recorder 6 ...
Страница 174: ...6 Data Recorder Documenting measurements 6 42 Edition 4 05 2014 DMS Go ...
Страница 175: ...DMS Go Edition 4 05 2014 7 1 Special functions of the DMS Go TC 7 ...
Страница 189: ...DMS Go Edition 4 05 2014 8 1 Maintenance and care 8 ...
Страница 195: ...DMS Go Edition 4 05 2014 9 1 Interfaces and Peripherals 9 ...
Страница 199: ...DMS Go Edition 4 05 2014 10 1 Appendix 10 ...
Страница 227: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 29 10 9 Recycling directives Overview 2 3 3 ...
Страница 228: ...10 Appendix Recycling directives 10 30 Edition 4 05 2014 DMS Go 4 5 1 4 ...
Страница 230: ...10 Appendix Recycling directives 10 32 Edition 4 05 2014 DMS Go Materials to be disposed of separately 1 ...
Страница 231: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 33 2 ...
Страница 233: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 35 Other materials and components 2 3 3 ...
Страница 234: ...10 Appendix Recycling directives 10 36 Edition 4 05 2014 DMS Go 1 3 ...
Страница 235: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 37 1 1 ...
Страница 236: ...10 Appendix Recycling directives 10 38 Edition 4 05 2014 DMS Go 5 ...
Страница 237: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 39 4 7 ...
Страница 238: ...10 Appendix Recycling directives 10 40 Edition 4 05 2014 DMS Go 3 6 6 6 7 ...
Страница 242: ...10 Appendix Recycling directives 10 44 Edition 4 05 2014 DMS Go ...
Страница 243: ...DMS Go Edition 4 05 2014 11 1 Specifications 11 ...
Страница 252: ...11 10 Edition 4 05 2014 DMS Go ...
Страница 253: ...DMS Go Edition 4 05 2014 12 1 Index 12 ...