GE DM5E Скачать руководство пользователя страница 1

GE

Sensing & Inspection Technologies

Ultrasonic

P/N 1259551 Rev. 1

February 2009

DM5E

Operating Manual

Содержание DM5E

Страница 1: ...GE Sensing Inspection Technologies Ultrasonic P N 1259551 Rev 1 February 2009 DM5E Operating Manual ...

Страница 2: ......

Страница 3: ...ESensingInspection com 2009 General Electric Company All rights reserved Technical content subject to change without notice DM5E Corrosion Thickness Gauge Operating Manual P N 1259551 Rev 1 February 2009 ...

Страница 4: ... no content intended for this page proceed to next page ...

Страница 5: ... the Instrument ON and OFF 3 1 3 Key Features of the DM5E 3 1 3 1 DM5E Corrosion Thickness Measurement Base Instrument 5 1 4 Contents of this Manual 6 Chapter 2 Understanding the Keypad Menu System Displays 2 1 Keypad Features 7 2 2 Interpreting the Display Screens 9 2 3 Working with the Configuration Display 14 ...

Страница 6: ...ss 25 3 3 Instrument Calibration 26 3 4 Setting the Maximum and Minimum Alarms 30 3 5 B SCAN Display Range 31 3 6 Locking and Unlocking Instrument the Controls 32 Chapter 4 Measuring Thickness 4 1 Selecting the Displayed View Mode 36 4 2 Normal Measurement View Mode 39 4 3 Min Scan and Max Scan Measurement View Modes 40 4 4 Differential Rate of Reduction Measurement View Mode 41 4 5 B SCAN Measure...

Страница 7: ...le 49 5 2 2 Erasing a Data Recorder File 50 5 3 Recording Thickness Measurements in Data Recorder Files 50 5 3 1 Storing Minimum of B Scan in Data Recorder Files 51 5 3 2 Navigating Through Data Recorder Files 51 Chapter 6 I O Features 6 1 Connecting the DM5E to a PC 54 Chapter 7 Specifications 7 1 Instrument Specifications 56 7 2 Data Recorder Option Features 60 7 3 DM5E Probe Transducer Specific...

Страница 8: ...1 Resetting the Operating Software 66 A 2 Firmware Updates 67 A 3 Web Site Information 69 A 4 EMC Documentation 70 A 5 Manufacturing Customer Service Locations 71 Appendix B Environmental Compliance B 1 Waste Electrical and Electronic Equipment WEEE Directive 74 B 2 Battery Disposal 75 B 2 1 What do the Markings Mean 75 B 2 2 The Risks and Your Role in Reducing Them 76 ...

Страница 9: ...s Proper use of ultrasonic test equipment requires three essential elements Selection of the correct test equipment Knowledge of the specific test application requirements Training on the part of the instrument operator This operating manual provides instruction in the basic setup and operation of the thickness gauge There are however additional factors which affect the use of ultrasonic test equi...

Страница 10: ...e required by a particular test Operators must understand Sound wave propagation theory Effects of the velocity of sound of the test material Behavior of the sound wave where two different materials are in contact Areas covered by the sound beam More specific information about operator training qualification certification and test specifications is available from various technical societies indust...

Страница 11: ...the condition of areas not inspected based on data from the evaluated areas should only be attempted by personnel fully trained in applicable statistical and probability techniques In particular materials subject to erosion or corrosion in which conditions can vary significantly in any given area should only be evaluated by fully trained and experienced operators Sound beams reflect from the first...

Страница 12: ...the values found in published tables In all cases best results are obtained if the instrument is calibrated on a velocity reference block made from the same material as the test piece this block should be flat and smooth and as thick as the maximum thickness of the test piece Operators should also be aware that the sound velocity may not be constant in the material being tested heat treating for e...

Страница 13: ... wear of the front surface Badly worn transducers will have a reduced effective measuring range The specified range of the transducer must include the complete range of thicknesses to be tested The temperature of the material to be tested must be within the transducer s temperature range 5 Use of Couplants Operators must be familiar with the use of ultrasonic couplants Testing skills must be devel...

Страница 14: ...cer may be a doubled reading and the thickness of the material being tested should be verified by the use of other methods If the transducer shows any sign of wear doubling may occur at a second echo or other echo signal combinations may produce a readable signal The instrument reading and apparent thickness are up to about twice the actual value resulting in a thickness greater than twice the min...

Страница 15: ...products recommended by us for battery operation Use of well known battery brands is highly recommended IMPORTANT Disposal of all components of the DM5E and its batteries must be done in compliance with all applicable regulations See Appendix B Environmetal Compliance for detailed information Software According to the current state of the art software is never completely free from errors Before us...

Страница 16: ...es The instrument no longer operates perfectly The instrument has been subjected to prolonged storage under adverse conditions such as exceptional temperatures and or especially high air humidity or corrosive environmental conditions The instrument has been subjected to heavy stresses during transportation Service Every effort has been made to provide you with a reliable product However should ser...

Страница 17: ... and the next chapter explains the keypad functions and display menu contents Closely reviewing the material in these two chapters will help you make better use of the more detailed information found in subsequent chapters of this manual 1 1 Supplying Power to the DM5E The instrument is powered by two AA size batteries providing up to approximately 100 hours of service life To install the batterie...

Страница 18: ...ual 1 1 Supplying Power to the DM5E cont Figure 1 DM5E Connections and Batteries shown with battery cover removed Rear of Instrument I O Device Connection mini USB Probe Cable Connection Install 2 AA batteries in compartment as shown Top of Instrument ...

Страница 19: ...ver your settings are saved and restored when the instrument is turned ON again When testing in remote locations always carry spare batteries 1 2 Powering the Instrument ON and OFF Power the instrument ON by pressing and holding until the instrument display activates Press and hold while the instrument is ON to turn it OFF 1 3 Key Features of the DM5E Large hollow filled thickness digits Alarm bar...

Страница 20: ...2 AA alkaline batteries Easy to use single level menu system Multiple display screen languages User selectable measurement resolution to 0 001 inch 0 01 mm Normal thickness MinScan MaxScan and Differential Rate of Reduction Measurement modes are standard Timed B Scan Optional Dual Multi function DM5E and DM5E DL Optional data recorder DM5E DL stores up to 50 000 thickness measurements in data reco...

Страница 21: ...E instrument 2 AA batteries Plastic carrying case Couplant sample Upgrade Utility Program on CD ROM requires mini USB to PC cable Operating manual in CD ROM Operating instruction card Certificate of conformity Note A mini USB to PC cable is only included with the DM5EDL version This cable can be purchased separately for use with the DM5E Basic or DM5E version ...

Страница 22: ...Manual This DM5E Operating Manual is divided into eight chapters and one appendix All chapters apply to all instrument versions except for two functions Data Recorder which is described in Chapter 5 applies only to the DM5E DL version Dual Multi mode applies only to the DM5E with Dual Multi functionality and to the DM5E DL Note Any DM5E can be retrofitted to incorporate the above functions ...

Страница 23: ...ted The display screens shown in this chapter are typical of those found in all the instrument configurations 2 1 Keypad Features The instrument s keypad includes a combination of dedicated keys and virtual function keys The key functions as illustrated in Figure 2 on page 8 are as follows Arrow Keys Use to adjust the value of the selected parameter and for on screen navigation Function Keys Use t...

Страница 24: ...ment display Press to activate the functional control when present above the key Press to launch the calibration process or make on screen selections Press and hold to power the instrument ON or OFF Press to adjust the value of the selected parameter and for on screen navigation Press to activate standard display modes shown here and the optional display modes if installed Optional display modes i...

Страница 25: ...nstalled Probe Setup Display Mode Allows the user to select a standard preloaded instrument setup each of which is matched to a specific probe model see Figure 4 on page 11 File Display Mode Allows the user to create and store thickness readings in data recorder files This feature is only available when the data recorder option is installed and activated see Figure 5 on page 12 When installed and ...

Страница 26: ...Chapter 2 Understanding the Keypad Menu System Displays 10 DM5E Operating Manual 2 2 Interpreting the Display Screens cont Figure 3 Measurement Display Mode ...

Страница 27: ...Chapter 2 Understanding the Keypad Menu System Displays DM5E Operating Manual 11 2 2 Interpreting the Display Screens cont Figure 4 Probe Setup Display Mode ...

Страница 28: ...Chapter 2 Understanding the Keypad Menu System Displays 12 DM5E Operating Manual 2 2 Interpreting the Display Screens cont Figure 5 File Display Mode ...

Страница 29: ...Chapter 2 Understanding the Keypad Menu System Displays DM5E Operating Manual 13 2 2 Interpreting the Display Screens cont Figure 6 Configuration Display Mode ...

Страница 30: ...file display MINIMUM ALARM Enables and sets the minimum alarm value see Section 3 4 on page 30 MAXIMUM ALARM Enables and sets the maximum alarm value see Section 3 4 on page 30 UNITS Sets the units of measurement to inches or millimeters see Section 3 2 on page 21 NOMINAL THICKNESS Sets the nominal thickness value that is then used to calculate and the differential and rate of reduction measuremen...

Страница 31: ...level LANGUAGE Sets the instrument display language see Section 3 2 on page 21 BSCAN MINIMUM Sets the minimum value that can be drawn when BSCAN is processing see Section 3 2 1 on page 23 BSCAN MAXIMUM Sets the maximum value that can be drawn when BSCAN is processing see Section 3 2 1 on page 23 DUAL MULTI Used to switch between Dual Multi and IP test modes when Dual Multi option is allowed see Se...

Страница 32: ... on page 21 GAIN Used to select a different level of gain LOW HIGH AUTO DEFAULT see Section 3 2 1 on page 23 RADIX Selects a period or comma to be used as a decimal point see Section 3 2 on page 21 AUTO POWER DOWN When set to ON this control conserves battery life by powering the instrument OFF if no key presses or measurements occur for the set number of minutes When set to OFF the instrument is ...

Страница 33: ...re automatically set to a default value based on the Setup file selected by the user Each instrument is supplied with standard Setup files and naming these to match the probe is the first step in setting up the instrument In preparation for measuring thickness this chapter explains how to Install a probe and configure the instrument by loading a Setup file see Section 3 1 on page 18 Set up the ins...

Страница 34: ...al 3 1 Connecting a Probe and Loading a Setup File Prior to measuring thickness you must connect a probe to the instrument see Figure 7 below and select a Setup file that is compatible with the probe Figure 7 Top View of Instrument Probe Cable Connection ...

Страница 35: ...nnecting a Probe and Loading a Setup File cont After a probe is connected press to activate the Probe Setup display The Probe Setup display which is shown in Figure 8 below allows the user to select one of the standard Setup files Figure 8 Selecting a Setup File ...

Страница 36: ...oading a Setup File cont The following instrument settings are automatically adjusted when the Setup file is activated View Mode Gain Measured Value display resolution Measurement Mode Nominal Thickness when Differential RR mode are selected Minimum Alarm Setting Maximum Alarm Setting ...

Страница 37: ...as necessary to access the Configuration display 2 Press or to select the desired parameter to modify 3 After the desired parameter is highlighted press to program that parameter 4 Press or to adjust the parameter value then press to confirm the new setting 5 Repeat steps 2 4 until all remaining parameters have been set After all of the settings have been configured press to return to the measurem...

Страница 38: ...al Multi Select ON or OFF Reading Hold Select HOLD or BLANK Time Set time in HH MM format Date Set date in Month Date Year format BSCAN MINIMUM BSCAN display of minimum value BSCAN MAXIMUM BSCAN display of maximum value Auto Power Down Select ON and then choose the desired time to automatically turn the instrument OFF 5 10 15 30 minutes after the last key press no data will be lost or select OFF t...

Страница 39: ...t gain level may set to AUTO DEFAULT HIGH or to a noise reducing LOW setting To select the instrument gain setting 1 Press 2 Select the GAIN control 3 Press to activate the control then use or to adjust the selected value and use or to select the cursor position 4 Press after the adjustment is complete ...

Страница 40: ...he Update Rate Measurements are updated at a rate of 4 or 8 Hz as selected by the user To select the measurement update rate 1 Press 2 Select the UPDATE RATE control 3 Press to activate the control then use or to adjust the setting 4 Press after the adjustment is complete ...

Страница 41: ...l values in the instrument s units and as a percentage that represent the variation from user specified nominal thickness see Section 4 4 To manually input or modify the nominal thickness 1 Press 2 Select the NOM THICKNESS control 3 Press to activate the control then use or to adjust the selected place value and or to select a different value 4 Press after the adjustment is complete ...

Страница 42: ...r 1pt 2pt or AUTO calibration mode in the LOCKOUT menu To complete the calibration follow the instructions on your instrument s display screen and in Figure 10 on page 27 1pt mode Figure 11 on page 28 2pt mode or Figure 12 on page 29 Auto mode Note Proper calibration is essential for accurate operation of the instrument The instrument should be calibrated if there is a change in the type or temper...

Страница 43: ...Chapter 3 Setting Up the DM5E DM5E Operating Manual 27 3 3 Instrument Calibration cont Figure 10 Instrument Calibration Procedure 1pt Mode ...

Страница 44: ...Chapter 3 Setting Up the DM5E 28 DM5E Operating Manual 3 3 Instrument Calibration cont Figure 11 Instrument Calibration Procedure 2pt Mode ...

Страница 45: ...Chapter 3 Setting Up the DM5E DM5E Operating Manual 29 3 3 Instrument Calibration cont Figure 12 Instrument Calibration Procedure Auto Mode ...

Страница 46: ...shes ON and OFF in 1 2 second intervals To input a minimum or maximum alarm thickness 1 Press to access the Configuration display see Figure 9 on page 21 2 Press or to select either MINIMUM ALARM or MAXIMUM ALARM 3 When the desired parameter is highlighted press to activate it 4 Press or to change the selected position in alarm thickness and then press or to adjust the value of the selected positi...

Страница 47: ... the preset range that point will be graphed as equal to the range limit 1 Press to access the Configuration display see Figure 9 on page 21 2 Press or to select either BSCAN MINIMUM or BSCAN MAXIMUM 3 When the desired parameter is highlighted press to activate it 4 Press or to change the selected position in alarm thickness and then press or to adjust the value of the selected position To switch ...

Страница 48: ...ver an upgrade of the basic version is possible by purchasing a license key In the Lockout menu the operator can set the probe zeroing and calibration mode CAL ZERO MODE by selecting 1pt 2pt or AUTO One of the following calibration reminder styles can also be selected REMIND TIME The calibration reminder is displayed after a preset elapsed time from the last thickness calibration The setting indic...

Страница 49: ...Chapter 3 Setting Up the DM5E DM5E Operating Manual 33 3 6 Locking and Unlocking Instrument the Controls cont Figure 13 Locking and Unlocking the Function Keys ...

Страница 50: ...Chapter 3 Setting Up the DM5E 34 DM5E Operating Manual no content intended for this page proceed to next page ...

Страница 51: ...ection 4 1 on page 36 Work with Normal measurement mode see Section 4 2 on page 39 Use the Min Scan and Max Scan measurement modes see Section 4 3 on page 40 Interpret the Differential Rate of Reduction measurement mode see Section 4 4 on page 41 Work with the B Scan measurement mode see Section 4 5 on page 42 Use the Dual Multi measurement mode see Section 4 6 on page 43 Note Load a Setup file co...

Страница 52: ...ed Recoupling the probe before the time out period expires allows the user to continue with the same evaluation period After the evaluation period is complete as indicated by the filling of the time out bar the minimum thickness observed during the evaluation period is displayed MAX SCAN Maximum Thickness Scan Allows the user to continuously evaluate material thickness provided the probe remains c...

Страница 53: ... minus nominal values expressed in percentage and dimensional terms appears in the middle of the display B SCAN B Scan Allows the user to continuously evaluate material thickness by graphing 1 point per second During the evaluation period the measured thickness value is displayed along with the minimum and maximum observed thickness values The display has a hidden 5 second time out bar which start...

Страница 54: ...surement mode displayed by the instrument may be modified as follows 1 Press as many times as necessary to access the Configuration menu 2 Press or to select the VIEW MODE parameter 3 Press to activate this parameter 4 Press or to scroll through the available modes then press to select the highlighted mode ...

Страница 55: ... view mode is active the display shows only a thickness reading as shown in Figure 14 below Note If the Data Recorder file is active the data location and stored value will be displayed at the top left corner of the display instead of the material velocity value Figure 14 NORMAL Measurement View Mode ...

Страница 56: ...he extreme minimum or maximum material thickness observed During the evaluation period thickness is displayed along with the minimum and maximum observed thickness values see Figure 15 below A time out feature allows a momentarily uncoupling of the probe during the evaluation period In these modes the instrument will continue to collect and compare readings for as long as the time out period is no...

Страница 57: ...ure 16 below displays both the currently measured thickness and a nominal thickness input by the user along with the difference between these two values measured minus nominal expressed in percentage and dimensional terms Note The displayed differences can have either positive or negative values Figure 16 DIFF RR Measurement View Mode ...

Страница 58: ...on period the thickness measurement is displayed along with the minimum and maximum thickness values A time out feature allows a momentarily uncoupling of the probe during the evaluation period In this mode the instrument will continue to collect and compare readings for as long as the time out period is not exceeded Pressing the SEND key will store the minimum value in the active file Figure 17 B...

Страница 59: ...ach probe will vary depending upon the type and thickness of the coating the thickness of the metal and the quality of the bond between the coating and the metal To activate or deactivate the Dual Multi mode you can either use the Configuration menu or press and hold for 1 5 seconds while in measurement mode A single peak icon above the battery icon indicates IP mode while a dual peak icon above t...

Страница 60: ...Chapter 4 Measuring Thickness 44 DM5E Operating Manual 4 6 Dual Multi Measurement View Mode cont Figure 18 DUAL MULTI Measurement View Mode Dual Multi mode deactivated Dual Multi mode activated ...

Страница 61: ... users to create Data Recorder files store thickness readings and the accompanying B SCAN when desired and navigate through file contents In this chapter you will learn how to Create Data Recorder files see Section 5 1 on page 46 Recall and erase Data Recorder files see Section 5 2 on page 49 Store thickness readings in Data Recorder files see Section 5 3 1 on page 51 Navigate Data Recorder file l...

Страница 62: ...the file naming process include The numbers of ROWS and COLUMNS in the Data Recorder file see Figure 20 on page 48 The ADVANCE DIRECTION see Figure 20 on page 48 NOTES related to the data file Note Once created the file name size and ADVANCE DIRECTION parameters cannot be edited Only the NOTES contents can be modified To accomplish this first select the file then activate and modify the parameter ...

Страница 63: ...Chapter 5 Using the Optional Data Recorder DM5E Operating Manual 47 5 1 Creating a New Data Recorder File cont Figure 19 Creating a New Data Recorder File ...

Страница 64: ...Chapter 5 Using the Optional Data Recorder 48 DM5E Operating Manual 5 1 Creating a New Data Recorder File cont Figure 20 Defining the Size and Advance Direction of a Data Recorder File ...

Страница 65: ...lable file locations but the advance direction cannot be changed 5 2 1 Recalling a Data Recorder File 1 Press as many times as necessary to access the Configuration display and then set DATA RECORDER to ON 2 Press again to launch the File Display mode 3 Activate the file selection function by pressing below FILES A file list will appear on the display 4 Press or to select the stored file you wish ...

Страница 66: ... cannot be retrieved 5 3 Recording Thickness Measurements in Data Recorder Files The Data Recorder must be enabled turned ON via the CONFIGURATION display Once the Data Recorder is turned on the contents of the active Data Recorder file are displayed in the File Navigation Grid shown in Figure 20 on page 48 To store a thickness reading in the selected file position press below SEND Remember that t...

Страница 67: ... is turned ON the contents of the active Data Recorder file are displayed in the File Navigation Grid shown in Figure 20 on page 48 After the navigation grid is activated and can be used to select any file position A heavier grid box appears around the selected file position Once a file position has been selected The current thickness reading can be sent to an EMPTY file position A measurement val...

Страница 68: ...Chapter 5 Using the Optional Data Recorder 52 DM5E Operating Manual no content intended for this page proceed to next page ...

Страница 69: ...ent The DM5E can communicate with external devices through its built in I O port A mini USB to PC cable is available for connection to this virtual COM port see Figure 21 below for the location With the UltraMATE UltraMATE Lite or even a simple Excel Macro a PC can import data recorder files from the instrument Figure 21 Location of Mini USB COM Port ...

Страница 70: ...ecting the DM5E to a PC The instrument s I O port can be connected to a PC using a mini USB cable Be sure to install the virtual COM port driver on the PC before making the connection Then follow the instructions on your PC screen to install the new hardware ...

Страница 71: ... 7 Specifications This chapter lists the features and specifications of your DM5E including Instrument specifications see Section 7 1 on page 56 Features of the optional Data Recorder see Section 7 2 on page 60 Probe specifications see Section 7 3 on page 61 ...

Страница 72: ...de 2 00mm to 127 0mm 0 079in to 5 00in in Dual Multi Measurement Mode the coating thickness range shall be 0 3mm to 2 50mm 0in to 0 098in Measuring Resolution 0 001 inch default selectable 0 001 0 01 inch 0 01 mm default selectable 0 01 0 1 mm Material Velocity Range 0 0200 to 0 7362 in msec 0 508 to 18 699 mm msec Material Velocity Resolution 0 0001 in msec 1 m s Units inch or millimeter Calibrat...

Страница 73: ...c Gain Control Display Type High resolution graphical LCD 64 x 128 Pixels 53 0mm x 27 0mm with backlight and adjustable contrast Update Rate 4 Hz or 8 Hz user selectable 24 Hz scan mode capture rate Thickness Value Display NORMAL view mode 5 digit 0 4 10 6 mm high B SCAN view mode 5 digit 0 1 2 55 mm high Display of Last Reading Solid filled or hollow digits indicate coupled or uncoupled condition...

Страница 74: ...r Requirements 2 AA size batteries Battery Life Operating Time Approximately 100 hours Instrument Shut 0ff Selectable ALWAYS ON or AUTO OFF after 5 10 15 30 minutes of inactivity Language Selectable English German French Spanish Italian Russian Japanese and Chinese I O Connectors Transducer Dual Lemo 00 coax Mini USB MINI B OTG to STD A Temperature Operating 10 F to 120 F 10 C to 50 C Storage 10 F...

Страница 75: ...ock IEC 68 2 27 Ea as per Mil Std 810C Method 516 2 Procedure I with a 15g 11ms impulse half sinusoidal wave applied 6 times per axis Sealing IEC529 IP54 Dust proof dripping water proof as per IEC 529 specifications for IP54 classification Note Instrument specifications are subject to change without prior notice ...

Страница 76: ...tures Capacity 50 000 readings File Structures Grid file Number of Rows 1 to 50 000 Number of Columns 1 to 223 File Naming Up to 24 character alphanumeric name Optional Software UltraMATE and UltraMATE Lite Note Instrument specifications are subject to change without prior notice ...

Страница 77: ...ingertip high temperature Operating Temperature Range continuous 10 to 70 C 10 to 70 C 10 to 70 C 10 to 204 C Contact Diameter 15 mm 590 20 mm 787 7 5 mm 300 12 7 mm 500 IP to First Measurement Range 1 0 to 200 mm 040 to 8 5 to 300 mm 200 to 12 0 6 to 60 mm 020 to 2 4 1 to 125 mm 20 C 040 to 5 68 F 1 3 to 25 4 mm 204 C 050 to 1 400 F Minimum Multi Echo Measurement Range 3 0 to 25 mm N A 2 0 to 10 ...

Страница 78: ...Chapter 7 Specifications 62 DM5E Operating Manual no content intended for this page proceed to next page ...

Страница 79: ...harsh solvents as they could cause the plastic to become brittle or damaged 8 2 Proper Cable Handling Avoid twisting or knotting of the cable Grasp the cable only by the connectors when connecting or disconnecting it from the instrument 8 3 Batteries Periodically inspect the batteries for signs of leakage or corrosion If either occurs remove and replace ALL the batteries Take care to properly disp...

Страница 80: ...Chapter 8 Maintenance 64 DM5E Operating Manual no content intended for this page proceed to next page ...

Страница 81: ...ry information including Resetting the instrument s operating software see Section A 1 on page 66 Upgrading the instrument s operating software see Section A 2 on page 67 Web site information see Section A 3 on page 69 EMC documentation see Section A 4 on page 70 Manufacturing and Service Center locations see Section A 5 on page 71 ...

Страница 82: ...oftware before resetting the operating software However any Data Recorder files stored in the instrument will not be deleted during this procedure To reset the instrument s operating software 1 Turn the instrument OFF 2 Press and hold Then press and hold for approximate 3 seconds until the display indicates that the reset is complete 3 A successful reset is acknowledged when the RESET COMPLETE mes...

Страница 83: ... button selecting PROGRAMS and then selecting the DM5E Upgrade Utility see Figure 22 on page 68 3 Select the Browse local disk button if the new firmware is stored on a local PC drive or select the Download new file button to connect to the GE Sensing Inspection Technologies FTP site you must be connected to the internet prior to this operation The program checks the firmware version you have on y...

Страница 84: ...Appendix A Supplemental Information 68 DM5E Operating Manual A 2 Firmware Updates cont Figure 22 Changing Parameter Settings ...

Страница 85: ...ANT All stored custom settings will be lost during the next step However stored Data Recorder files will not be affected 7 Reset the operating software to purge the instrument s memory A 3 Web Site Information The following information is available on the GE Sensing Inspection Technologies web site at www geinspectiontechnologies com Up to date information on operating software Latest upgrade util...

Страница 86: ...transducers operating at frequencies from 5 to 15 MHz This instrument displays a thickness calculated by the measured time of flight in the material at the velocity programmed into the instrument The instrument is intended for use as a portable thickness gauge in an industrial environment The instrument is not intended to be part of a closed control loop in an automated process control system Comp...

Страница 87: ...optimum quality of instrument conformance The DM5E is manufactured by GE Sensing Inspection Technologies China Q Section 4th Floor 185 Tai Gu Road Waigaoqiao Free Trade Zone Shanghai China Should you experience a problem or require technical assistance visit the product service page accessible at www geinspectiontechnologies com to locate your local GE Inspection Technologies representative or con...

Страница 88: ...Appendix A Supplemental Information 72 DM5E Operating Manual no content intended for this page proceed to next page ...

Страница 89: ...ntal Compliance DM5E Operating Manual 73 Appendix B Environmental Compliance This appendix contains information on the following topics WEEE Directive see Section B 1 on page 74 Battery disposal see Section B 2 on page 75 ...

Страница 90: ...ld impact health and the environment In order to avoid the dissemination of those substances in our environment and to diminish the pressure on the natural resources we encourage you to use the appropriate take back systems Those systems will reuse or recycle most of the materials of your end life equipment in a sound way The crossed out wheeled bin symbol invites you to use those systems If you n...

Страница 91: ...clude lettering to indicate cadmium Cd lead Pb or mercury Hg For proper recycling return the battery to your supplier or to a designated collection point B 2 1 What do the Markings Mean Batteries and accumulators must be marked either on the battery or accumulator or on its packaging depending on size with the separate collection symbol In addition the marking must include the chemical symbols of ...

Страница 92: ...nce toxic metals Pb for lead Hg for mercury and Cd for cadmium Cadmium poisoning can result in cancer of the lungs and prostate gland Chronic effects include kidney damage pulmonary emphysema and bone diseases such as osteomalcia and osteoporosis Cadmium may also cause anemia discoloration of the teeth and loss of smell anosmia Lead is poisonous in all forms It accumulates in the body so each expo...

Страница 93: ...enance 63 Operating Life 58 Size 58 B Scan Display Range Setting 31 Measurement Mode 42 Storing Minimum in Data Recorder File 51 C Cables Connections 2 I O Port 53 Maintenance 63 Calibration Procedure 1pt 27 Procedure 2 pt 28 Procedure Auto 29 When to Perform 26 COM Port see I O Port Configuration Adjusting Settings 21 Display Mode 13 Display Mode Settings 14 Menu 21 Parameters 22 ...

Страница 94: ...ng Erasing Files 49 Specifications 60 Storing B Scan Minimum 51 Storing Measurements 50 Date Manual i Setting 22 Differential Rate of Reduction Measurement Mode 41 Display Mode Configuration 13 File 12 Measurement 10 Options 9 Probe 11 Disposal Batteries 75 Electronic Waste 74 DM5E Key Features 3 Keypad 7 Power Required 1 Powering ON OFF 3 Setting Up 17 Versions Available 1 6 Dual Multi Measuremen...

Страница 95: ...ion Keys Locking Unlocking 33 G Gain Setting 23 I I O Port Cable 53 Type and Location 53 K Keypad Keys Description 7 Photograph 8 L Language Options 58 Setting 22 Locking Unlocking the Controls 32 M Maintenance Batteries 63 Cables 63 Instrument 63 Manufacturing Location 71 Measurement Mode B Scan 42 Differential Rate of Reduction 41 Dual Multi 43 Min Scan Max Scan 40 Normal 39 Options 36 Selecting...

Страница 96: ...rement Mode 39 P Part Number Manual i PC Connecting to 54 Power for DM5E 1 Powering ON OFF 3 Probe Connecting 2 18 Display Mode 11 Setup Display 19 Specifications 61 R Resetting the Operating Software 66 S Safety Information xiii Setup Display Probe 19 Setup File Loading 18 Parameters 20 Size Instrument 59 Software Resetting 66 Specifications Data Recorder 60 Instrument 56 Probe Transducer 61 T Te...

Страница 97: ...iii Transducer see Probe U Units Setting 22 Update Rate Setting 24 Updating the Firmware 67 Upgrade Utility Program 67 V Versions Available 6 View Mode see Measurement Mode W Warnings General vii Waste Disposal Batteries 75 Electronic Equipment 74 Web Site Information 69 WEEE Directive 74 Weight Instrument 58 ...

Страница 98: ...Index 82 DM5E Operating Manual no content intended for this page proceed to next page ...

Страница 99: ......

Страница 100: ...nue Canley Coventry CV4 8AF England Tel 44 845 130 3925 France 68 Chemin des Ormeaux Limonest 69760 France Tel 33 47 217 9216 Germany Robert Bosch Str 50354 Hürth Germany Tel 49 2233 601 0 Spain San Maximo 31 Planta 4A Nave 6 Madrid 28041 Spain Tel 34 195 005 990 China 5F Building 1 No 1 Huatuo Road Zhangjiang High Tech Park Shanghai 201203 China Tel 86 800 915 9966 toll free 86 0 21 3877 7888 Jap...