Equipment Description and General Construction Requirements
2.3 Room Structural Requirements
2.3.1 Floor Requirements
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
2-40
t
Table 2-4: Floor Leveling Specifications
Item
Requirement
Floor leveling area
640 cm×370 cm (21'×12') (covering the entire planned area of table and gantry installation,
depending on room layout)
Slope
3 mm (0.125") over 3048 mm (120")
Floor surface
A single poured surface
Flatness
The surface must be smooth and without significant valleys or peaks.
The entire surface area must have an overall flatness of 5 mm over 1500 mm in any direction
(see
App.B, Measuring Floor Flatness
for measurement procedure).