Equipment Description and General Construction Requirements
2.2 Room Size, Layout and Considerations
2.2.2 System Layout Drawings
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
2-18
Figure 2-11: Recommended Layout
Note
:
Dimensions are in mm
Le
ad gl
ass
w
ind
ow
O Extender Layout
350
ext
50
1600 – CT Scan Length
1150 – NMFOV to CT
50+100 400 FOV
6450 - Footprint
Pivot – 2000
250
420
0
6750
Defines the recommended
room dimensions required to
enable installation, operation
and service of the system in
safe conditions.
Does not take into account
local requirements.
Enables relative freedom of
movement for operator and
service.
Enables installation of table
extender.
Can keep two collimator carts
in the room.
The operator room is not taken
into account in the room
dimensions.
Space must be allocated for
the operator room, but the
location can differ.