Electrical Requirements
5.5 System Cable Information
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
5-12
5.5 System Cable Information
This section provides technical information regarding system cables connecting different
subsystems, in order to facilitate the planning of cable routing.
Table 5-5: D670 Sub-system Inter-connection Cables
Start / Destination
H/V
Separation
(Y or N)
Run
#
Actual Length
(Usable Length)
*
Description
From
To
Feet
Meters
PDU
CT gantry
Y
050
63 (55)
19.3 (16.76)
HVDC, PDU to gantry
PDU
CT gantry
Y
051
62.5 (55)
19 (16.76)
HVAC, PDU to gantry
PDU
CT gantry
Y
052
63 (57)
19.3 (17.56)
LVAC, PDU to gantry
PDU
Console
(adaptor box)
Y
053
80 (75)
24.5 (22.86)
LVAC, PDU to console
PDU
CT gantry
N
055
63 (55)
19.3 (16.76)
Ground, PDU to raceway
Console
CT gantry
N
056
83 (75)
25.5 (22.86)
Ground, raceway to console
PDU
CT gantry
N
100
70(62)
21.4 (18.86)
Signal, gantry MSUB to PDU
Console
CT gantry
N
101
86(78)
26.35 (23.71) Signal, gantry MSUB to OC
Console
CT gantry
N
102
86 (81)
26.3 (24.86)
Signal (LAN), gantry to OC
Console
CT gantry
N
103
80 (75)
24.3 (22.86)
Fiber optic, gantry to OC
PDU
UPS
Y
15 (12)
4.6 (3.6)
Power cable, CT PDU to UPS
PDU
UPS
Y
15 (12)
4.6 (3.6)
Power cable, UPS disconnect panel to CT PDU
UPS
MDP (A1)
†
Y
45 (37)
13.6 (11.3)
UPS control cable
Injector
CT gantry
N
100
30.5
Gantry to injector