Chapter 12 - Troubleshooting
Page 333
12.4.2 Base Axis Does Not Move
If the base axis does not move, perform the following checks:
Figure 12-12:
Troubleshooting for Base Axis Does Not Move
Base Axis does not move
Is E-STOP circuitry
activated or Cradle Release
button pressed?
Reset the E-STOP
circuitry or press the
Cradle Release button
yes
no
Does a mechanical interference exist?
Check the locking mechanism including the
solenoid functionality, use the PET position Jog
button on the Control Panel
yes
no
Release the
mechanical
interference
Check:
1. The path from the ELEVATION/TILT(BASE Axis ) Amplifier to
the motor include the Amplifier itself
PNs:
1. 2323724
2. 2309058
3. 2323720
4. 2323722
End of test
Содержание Discovery LS
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION 2307224 100 Rev 6 Discovery LS System Service Manual ...
Страница 2: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 2 ...
Страница 24: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 24 Table of Contents ...
Страница 32: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 32 List of Figures ...
Страница 110: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 110 Chapter 5 Console ...
Страница 222: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 222 Chapter 6 Table ...
Страница 242: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 242 Chapter 7 Gantry ...
Страница 357: ......