Chapter 1. Programming Site Data
16
DigitalFlow™ XGS868i Programming Manual
1.8.2 Initializing Setup Parameters - Default Setup
Use this option to initialize (reset) all of the parameters within the
Setup
menu to their default values. Complete the
following steps to reset all of the parameters:
1.
In the
Keypad Program
, scroll to
PROG
and press
[Enter]
.
2.
In the
PROG
menu, scroll to
CH1
or
CH2
and press
[Enter]
.
3.
In the
Channel PROGRAM
menu, scroll to
Setup
and press
[Enter]
.
4.
Scroll to
INIT
. Press
[Enter]
.
5.
Scroll to
Yes
to initialize default parameters or
No
to abort this command. Press
[Enter]
.
Procedure Options
After completing the above steps, the XGS868i returns to the
Channel Set up
window. Do one of the following:
•
To continue programming, refer to Appendix A,
Menu Maps
, to navigate to the desired menu.
•
To leave the
Keypad Program
, press
[Escape]
four times.
M>S Switch
0 to 250
μ
sec
50
μ
sec
If the burst mode is set to
Skan/Measure
(S/M), the meter switches from
Skan
to
Measure Mode
when Delta T is less than the M>S_Switch value.
DO NOT
change this value unless advised by the factory.
# Shifts
0 to 10
3
The number of shifts corresponds to the actual number of transmits per
cycle (number of signals added together in one direction to produce an
averaged signal for one interrogation of the fluid) and need only be changed
if the environment is very noisy or the acoustic signal is weak.
A Divisor
0.1 to 10
2.5
The Divisor used to calculate the Measure Mode integrated threshold level is
not normally changed.
# Transmit
Pulses
1 to 16
4
# Transmit Pulses specifies the number of pulses in a burst. For difficult con-
ditions (i.e. long paths, high velocity or high temperature), settings as high as
16 may be necessary.
T Window
(cycles)
0 to 1000
0
The XGS868 calculates the size of the transmit window based on pipe size
and fluid sound speed. However, for special diagnostic purposes, it is
possible to reset the window size.
R Window
(cycles)
10 to 128
10
The XGS868 calculates the size of the receive window based on pipe size and
fluid sound speed. However, for special diagnostic purposes, it is possible to
reset the window size.
Table 4: Transducer Signal Settings (cont.)
Transducer
Signal
Parameters
Range
Default
Value
Miscellaneous Information
Содержание DigitalFlow XGS868i
Страница 2: ......
Страница 4: ... no content intended for this page ...
Страница 10: ...Contents viii DigitalFlow XGS868i Programming Manual ...
Страница 52: ...Chapter 1 Programming Site Data 42 DigitalFlow XGS868i Programming Manual ...
Страница 64: ...Chapter 2 Displaying Data 54 DigitalFlow XGS868i Programming Manual no content intended for this page ...
Страница 78: ...Chapter 3 Logging Data 68 DigitalFlow XGS868i Programming Manual no content intended for this page ...
Страница 80: ...Chapter 4 Printing Data 70 DigitalFlow XGS868i Programming Manual no content intended for this page ...
Страница 84: ...Chapter 5 Clearing Data 74 DigitalFlow XGS868i Programming Manual no content intended for this page ...
Страница 85: ...DigitalFlow XGS868i Programming Manual 75 Appendix A Menu Maps Appendix A Menu Maps ...
Страница 86: ...76 DigitalFlow XGS868i Programming Manual no content intended for this page ...
Страница 96: ...86 DigitalFlow XGS868i Programming Manual no content intended for this page ...
Страница 146: ...Appendix C Programming the XGS868i Using PanaView 136 DigitalFlow XGS868i Programming Manual ...
Страница 148: ...138 DigitalFlow XGS868i Programming Manual no content intended for this page ...
Страница 174: ...Warranty 164 DigitalFlow XGS868i Programming Manual no content intended for this page ...
Страница 175: ......