DigitalFlow™ XGM868i Startup Guide
61
Appendix A. CE Mark Compliance
Appendix A. CE Mark Compliance
A.1 Introduction
For CE Mark compliance, the Model XGM868i flow transmitter must be wired in accordance with the instructions in
this appendix.
IMPORTANT:
CE Mark compliance is required only for units intended for use in EEC countries.
A.2 Wiring
The Model XGM868i must be wired with the recommended cable, and all connections must be properly shielded and
grounded. Refer to Table 13 for the specific requirements.
Note:
If the Model XGM868i is wired as described in this appendix, the unit will comply with the EMC Directive
2004/108/EC.
Table 13: Wiring Requirements
Connection
Cable Type
Ground Termination
Transducer
Armored RG62 a/U
Grounded via cable gland.
Input/Output Armored 22 AWG shielded (e.g. Baystate #78-1197)
with armored material added to outside of jacket
Grounded via cable gland.
Power
Armored 14 AWG 3 conductor
Grounded via cable gland.
Содержание DigitalFlow XGM868i
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 30: ...Chapter 1 Installation 22 DigitalFlow XGM868i Startup Guide no content intended for this page ...
Страница 48: ...Chapter 2 Initial Setup 40 DigitalFlow XGM868i Startup Guide no content intended for this page ...
Страница 50: ...Chapter 2 Initial Setup DigitalFlow XGM868i Startup Guide 42 no content intended for this page ...
Страница 68: ...Chapter 4 Specifications 60 DigitalFlow XGM868i Startup Guide no content intended for this page ...
Страница 70: ...Appendix A CE Mark Compliance 62 DigitalFlow XGM868i Startup Guide no content intended for this page ...
Страница 76: ...Appendix B Data Records 68 DigitalFlow XGM868i Startup Guide no content intended for this page ...
Страница 84: ...Warranty 76 DigitalFlow XGM868i Startup Guide no content intended for this page ...
Страница 86: ......
Страница 88: ......
Страница 90: ......
Страница 91: ......