Programming Site Data
1-37
March 2008
Setting Up Analog Outputs
1.
At the
Slot X Outputs
prompt, press
[F1]
to set up output A or
[F2]
to set up output B.
Note:
The set up of output A is used as an example. Identical
procedures would be used to set up output B.
2.
Press
[F1] = OFF
to disable output A and return to the previous
prompt, or press
[F2] = 0-20 m or [F3] = 4-20 m
to specify the
desired range for output A.
3.
Use the [
←]
and [
→]
and
[F1]-[F4]
keys to select a
Channel
option
for the data source.
4.
Use the [
←]
and [
→]
and
[F1]-[F4]
keys to specify the desired
Output Measurement
parameter.
5.
Enter the
Zero
value for the low end of the chosen output range.
Press
[ENT]
.
6.
Enter the
Full
value for the high end of the chosen output range.
Press
[ENT]
.
7.
Press
[F1]
-
[F2]
to set up another output or press
[EXIT]
to select
another slot for set up.