April 2008
2-2
Error Codes and Screen Messages
E0: No Error
Problem:
No error condition currently exists.
Cause:
This message appears briefly to confirm that the cause
for a previous error condition has disappeared, has
cleared or been resolved.
Action:
No action is required.
E1: Low Signal
Problem:
Poor ultrasonic signal strength or the signal exceeds the
limits entered via the
User Program
.
Cause:
Poor signal strength may be caused by a defective cable,
preamplifier or decoupler, a flowcell problem, a defective
transducer or a problem in the electronics console. A
signal that exceeds the programmed limits is probably
caused by the entry of an improper value in the
PIPE
or
SETUP SIGNL
submenus of the
User Program
.
Action:
Using the procedures in Chapter 3,
Diagnostics
, check
the components listed above. Also, check the value
programmed into the
PIPE
or
SETUP SIGNL
submenu, as
described in the
Operator’s Manual
.
E2: Sound Speed Error
Problem:
The sound speed exceeds the limits programmed in the
SETUP SIGNL
submenu of the
User Program
.
Cause:
The error may be caused by incorrect programming, poor
flow conditions, wrong transducer spacing or poor
coupling.
Action:
Compare the measured sound speed to tabulated nominal
values for the fluid being used and correct any
programming errors. Refer to Chapter 3,
Diagnostics
, to
correct any flowcell and/or transducer problems.
E3: Velocity Range
Problem:
The velocity exceeds the limits programmed in the
SETUP
submenu of the
User Program
.
Cause:
This error may be caused by the entry of improper
programming data or by poor flow conditions and/or
excessive turbulence.
Action:
Make sure the actual flow rate is within the programmed
limits. See the
Programming Manual
for details. Refer to
Chapter 3,
Diagnostics
, to correct any flowcell and/or
transducer problems.
Содержание DigitalFlow GC868
Страница 6: ...Chapter 1 ...
Страница 23: ...Chapter 2 ...
Страница 31: ...Chapter 3 ...
Страница 39: ...Chapter 4 ...
Страница 52: ...Appendix A ...
Страница 53: ...Service Record Introduction A 1 Data Entry A 1 Diagnostic Parameters A 3 ...
Страница 55: ...April 2008 A 2 Service Record Table A 1 Service Record Continued Date Description of Service Performed By ...
Страница 57: ...Appendix B ...