April 2008
1-10
Calibration
Numer Option = [F2]
The Numer option allows the user to calculate the temperature vs.
resistance information. To do this, the user must first collect data
using the RTDs option in the PRINT menu (discussed in Chapter 5 of
the
Programming Manual
). The RTDs option allows you to know
what point value the GC868 is reading at a given temperature. With
this information, the Numer option enables you to enter the Set
Temperature and the slope of the RTD input in points/degree,
allowing absolute control over RTD calibration. Before attempting
this step, follow the instructions in Chapter 5 of the
Programming
Manual
(“The RTDs Submenu”) to obtain point values at your set
point and slope point.
1.
Press
[F1]
to select Input A, or
[F2]
to select Input B.
Set Number
2.
Press
[F1]
to enter the set number (formerly known as the zero
point).
3.
Enter the set point temperature, and press
[ENT].
4.
The program then asks for the set point number. Enter the points
measured at the set temperature, and press
[ENT].
Slope Number
5.
Press
[F2]
to enter the slope number. Calculate the slope number
with the formula:
(1-1)
6.
Enter the RTD slope number and press
[ENT].
Procedure Options
You have completed calibration of the RTD input. Do one of the
following:
•
To calibrate another input, press
[EXIT].
•
To calibrate additional slots, press
[EXIT]
and then
[F4] (EXIT)
and
proceed to the appropriate section.
•
To leave the Calibration Menu, press
[F4] (EXIT)
twice.
Slope
Slope point
set point
–
slope temp
set temp
–
------------------------------------------------------
=
Содержание DigitalFlow GC868
Страница 6: ...Chapter 1 ...
Страница 23: ...Chapter 2 ...
Страница 31: ...Chapter 3 ...
Страница 39: ...Chapter 4 ...
Страница 52: ...Appendix A ...
Страница 53: ...Service Record Introduction A 1 Data Entry A 1 Diagnostic Parameters A 3 ...
Страница 55: ...April 2008 A 2 Service Record Table A 1 Service Record Continued Date Description of Service Performed By ...
Страница 57: ...Appendix B ...