Diagnostics
3-7
March 2008
Transducer Problems
Ultrasonic transducers are rugged, reliable devices. However, they are
subject to physical damage from mishandling and chemical attack.
Clamp-on transducers are also subject to installation variables such as
physical misalignment and faulty coupling to the pipe on which they
are mounted.
Because, transducer problems are largely dependent on the type of
transducers used, wetted or clamp-on, the following list of potential
problems is grouped according to transducer type. Contact the factory
if you cannot solve a transducer-related problem.
Wetted Transducer
Problems
1.
LEAKS:
Leaks may occur around the transducer and/or the flowcell
fittings. Repair such leaks immediately. If the leaking fluid is
corrosive, carefully check the transducer and cables for damage,
after the leak has been repaired.
2.
CORROSION DAMAGE:
If the transducer material was not
properly chosen for the intended application, the transducers may
suffer corrosion damage. The damage usually occurs either at the
electrical connector or on the transducer face. If corrosion is
suspected, remove the transducer from the flowcell and carefully
inspect the electrical connector and the transducer face for
roughness and/or pitting. Any transducer damaged in this manner
must be replaced. Contact the factory for information on
transducers in materials suitable for the application.
3.
INTERNAL DAMAGE:
An ultrasonic transducer consists of a
ceramic crystal bonded to the transducer case. The bond between
the crystal and the case, or the crystal itself, may be damaged by
extreme mechanical shock and/or temperature extremes. Also, the
internal wiring can be corroded or shorted if contaminants enter
the transducer housing.
4.
PHYSICAL DAMAGE:
Transducers may be physically damaged by
dropping them onto a hard surface or striking them against another
object. The transducer connector is the most fragile part and is the
one most subject to damage. Minor damage may be repaired by
carefully bending the connector back into shape. If the connector
cannot be repaired, the transducers must be replaced.
IMPORTANT:
Transducers must be replaced in pairs. Refer to
Chapter 2,
Initial Setup
, of the
Startup Guide
to
program the replacement transducer data into the
meter.
If the instructions in this section fail to resolve the problem, contact
the factory for assistance.
Содержание DigitalFlow DF968
Страница 2: ......
Страница 7: ...Chapter 1 ...
Страница 23: ...Chapter 2 ...
Страница 32: ...Chapter 3 ...
Страница 42: ...Chapter 4 ...
Страница 55: ...Appendix A ...
Страница 56: ...Service Record Introduction A 1 Data Entry A 1 ...
Страница 58: ...March 2008 A 2 Service Record Table A 1 Service Record Continued Date Description of Service Performed By ...
Страница 60: ...Appendix B ...
Страница 72: ...March 2008 Optional Enclosures B 11 Figure B 2 Model DF868 Rack Mount Console Assembly ref dwg 705 849 ...