DigitalFlow™ Multipurpose Ultrasonic Liquid Flowmeter Startup Guide (1- and 2-Channel)
19
Chapter 2. Initial Setup
2.4 Activating a Channel
The
ACTIV
submenu permits selection of the desired measurement method. In addition, it is used to activate/deactivate
one or both of the channels in a 2-Channel Model DF868.
2.4.1 1-Channel Meter
1.
Enter the
ACTIV
submenu by pressing
[F1]
at the
User PROGRAM
prompt.
2.
Press
[F1]
to activate the channel in
Transit-Time
mode,
or
[F2]
to activate the channel in
TransFlection
mode.
2.4.2 2-Channel Meter
1.
Enter the
ACTIV
submenu by pressing
[F1]
at the
Channel PROGRAM
prompt.
2.
Press
[F1]
(Off) to deactivate the channel and return to the
Channel PROGRAM
prompt, or press
[F2]
to activate the
channel in
Transit-Time
mode, or
[F3]
to activate
TransFlection
mode.
Note:
If a customer has not ordered the DF868 with TransFlection mode, the
TRNFL
option does not appear.
Proceed directly to the next section to program the
SYSTM
submenu.
2.5 Entering System Data for a Channel
Begin the programming of the
SYSTM
submenu in either the
1-Channel
or
2-Channel
section below.
2.5.1 1-Channel Meter
For the 1-Channel Model DF868, the information entered in the
SYSTM
submenu pertains to the global operation of the
flowmeter.
1.
At the
User Program
screen, press the
[F2]
function key to program the
SYSTM
submenu.
2.
Enter a
Site Label
of up to 9 characters and press the
[ENT]
key. (While taking measurements, the site label appears
on the locator bar.)
3.
Enter a
Site Message
of up to 21 characters. Use this screen to enter a brief description of the site. When the
message has been entered, press the
[ENT]
key.
4.
Press
[F1]
to turn the
Energy Option
OFF
or press
[F2]
to turn it
ON
. (The Energy Option calculates the energy of a
system based on temperatures at the supply and return points, and the flow of fluid.)
5.
To select the
System Units
, press
[F1]
to display parameters and measurements in English units or press
[F2]
to
display parameters and measurements in Metric units.
6.
To program the
Totalizer Option,
press
[F1]
to measure totals automatically, or press
[F2]
to measure totals
manually.
The remainder of the
SYSTM
submenu is identical for the 1-Channel and 2-Channel versions of the Model DF868.
Proceed to the
1- and 2-Channel Meters
section to complete the programming of this submenu.