Chapter 1. Calibration
10
DigitalFlow™ Multipurpose Ultrasonic Liquid Flowmeter Service Manual
1.6.2 Testing the Alarm Relays
Note:
The procedure for testing
Alarm Relay A
is identical to that for testing
Alarm Relays B and C
. However, make sure
that the ohmmeter is connected to the desired normally-open or normally-closed contact of the currently
selected relay. See Figure 4 on page 9 for the correct pin numbers on the option card terminal block.
4.
Pressing
[F1] (CLOSE)
should yield an ohmmeter reading of about zero. Pressing
[F2] (OPEN)
should yield an infinite
ohmmeter reading.
5.
Press
[F3]
to
EXIT
.
6.
Repeat the above procedure until both the normally-open and normally-closed contacts for all three alarm relays
have been tested. Then, proceed to
Procedure Options
below.
Procedure Options
You have completed testing the alarm relays. Do one of the following:
•
To calibrate/test additional slots, press
[F4]
(
EXIT)
and proceed to the appropriate section.
•
To leave the
Calibration Menu,
press
[EXIT]
twice
.