Chapter 2. Error Codes
14
DigitalFlow™ XGM868i Service Manual
E0: No Error
Problem:
No error condition currently exists.
Cause:
This message appears briefly to confirm that the response to another error message has corrected the
problem.
Action:
No action is required.
E1: Low Signal
Problem:
Poor ultrasonic signal strength or the signal exceeds the limits entered via the
Keypad Program
.
Cause:
Poor signal strength may be caused by a defective cable, a flowcell problem, a defective transducer or a
problem in the electronics console. A signal that exceeds the programmed limits is probably caused by the
entry of an improper value in the
Channelx-Set up-Signal
submenu of the
Keypad Program
.
Action:
Using the procedures in Chapter 3,
Diagnostics
, check the components listed above. Also, check the limit
values entered into the
Channelx-Set up-Signal
submenu, as described in Chapter 1,
Programming Site
Data
, of the
Programming Manual
.
E2: Sound Speed Error
Problem:
The sound speed exceeds the limits programmed in the
Channelx-Set up-Signal
submenu of the
Keypad
Program
.
Cause:
The error may be caused by incorrect programming, poor flow conditions or poor transducer orientation.
Action:
Compare the measured sound speed to tabulated nominal values for the gas being used and correct any
programming errors. Refer to Chapter 3,
Diagnostics
, to correct any flowcell and/or transducer problems.
E3: Velocity Range
Problem:
The velocity exceeds the limits programmed in the
Channelx-Set up-Signal
submenu of the
Keypad
Program
.
Cause:
This error may be caused by the entry of improper programming data or by poor flow conditions and/or
excessive turbulence.
Action:
Make sure the actual flow rate is within ±75 ft/sec (±23 m/sec). See Chapter 1,
Programming Site Data
, in
the
Programming Manual
for details. Refer to Chapter 3,
Diagnostics
, to correct any flowcell and/or
transducer problems.
Содержание DigitalFlo XGM868i
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 44: ...Chapter 4 Parts Replacement DigitalFlow XGM868i Service Manual 36 ...
Страница 48: ...40 DigitalFlow XGM868i Service Manual no content intended for this page ...
Страница 70: ...62 DigitalFlow XGM868i Service Manual no content intended for this page ...
Страница 78: ...Warranty 70 DigitalFlow XGM868i Service Manual no content intended for this page ...
Страница 79: ......