Chapter 3: Hardware Overview
Definium AMX 700 X-Ray System
5161515-1EN
Rev. 6 (10 February 2008)
3-9
Operator Manual
© 2008 General Electric Company. All rights reserved.
NOTE:
The Collimator Lamp automatically deactivates upon prep. The Prep Cycle automatically drops out
and returns to standby if an exposure is not initiated within ten (10) seconds, or if the collimator
lamp is turned on. The exposure is automatically discontinued if the Prep/Expose button is released
before exposure is complete.
Test the Battery
Activate the TechSwitch transmitter for five (5) seconds and observe the Low Battery Indicator. Should
the indicator illuminate, replace the battery.
To test the additional spare TechSwitch transmitter: Set-up the X-ray machine to recharge the batteries.
Observe the Green 'ON' LED on the internal TechSwitch PCB Electronics Package. With the Transmitter
Unit in hand step back at least 24 feet (7 meters) from the X-ray unit, and cycle the TechSwitch system,
observing the indicator lamps on the top panel of the PCB.
NOTE:
No rotor boost or exposures will be made in this mode. Check to see that the TechSwitch system
functions reliably at this range. If not, a new battery is required. This test should be performed
quarterly or when a transmitter battery problem is suspected.
Replace Transmitter Battery
Remove the six screws from the gray back cover of the TechSwitch transmitter. Place a new 9-volt
alkaline battery in the direction shown on the inside of the housing. Replace the back cover securing the
six screws.
NOTE:
The battery should be replaced every two years under normal usage to ensure uninterrupted
operation.
Clean Transmitter Cradle
The TechSwitch remote Transmitter Unit and Transmitter Cradle have durable plastic casings that should
retain their appearance for many years. Both should be periodically cleaned with a soft cloth slightly
dampened with water and a mild soap or detergent.
Options and Accessories
Dose Area Product (DAP) Meter
Full field ion chamber with an integral display unit. The display unit contains electronics controlling total
dose area product, system reset and system test.
• Provides easy to read real time information on patient dose.
• Allows for simple manual transcription of this information into patient records.
FOR
TRAINING
PURPOSES
ONLY!
NOTE:
Once
downloaded,
this
document
is
UNCONTROLLED,
and
therefore
may
not
be
the
latest
revision.
Always
confirm
revision
status
against
a
validated
source
(ie
CDL).