3.4 Internal battery
WARNING
The S/5 Compact Monitor is always energized by an internal battery. A
short circuit may cause internal damage. Do not touch any exposed wiring
or conductive surface inside, as this may cause an electric shock.
The monitor can be run either on mains power or on internal battery. Battery operation is
initiated when the power cord is disconnected or when the mains power is lost during
monitoring.
The internal battery capacity is up to 1.5 hours with fully charged batteries.
NOTE: Before using the monitor for the first time, charge the battery to its full capacity
(charging time 10 hours).
3.4.1 Battery time
A symbol on the monitor screen indicates the remaining operating time of the battery.
When about 10 minutes of battery operating time is left, a 'Batt low' alarm is activated and the
monitor makes one beep.
When about two minutes of operating time is left, a new alarm 'Batt empty' is activated.
3.4.2 Refreshing the batteries
The battery is automatically charged when the monitor is connected to mains power. Charging
the battery depends on the discharge level of the battery. Battery charging time is ten hours to
full capacity.
If the monitor is used mainly with the mains power, refresh the battery by charging and
discharging the battery completely once a month to slow down loss of battery capacity.
NOTE: If the monitor is used for frequent short transportation, refresh the batteries every two
weeks and replace the batteries every six months.
To refresh:
1.
Connect the monitor power cord to power outlet for at least ten hours.
2.
Remove the power cord, and turn on the monitor and let it turn off by itself (it takes about
1.5 hours). Wait for 15 minutes.
3.
Reconnect the power cord to power outlet and charge the battery for at least ten hours.
NOTE: Inadequate maintenance will decrease the battery capacity and the monitor's operation
time on batteries.
When charging, a charging symbol is displayed.
During battery charging, the standby led is blinking. It lights up when the batteries are fully
charged.
3.4.3 Replacing the batteries
1.
Open the Bottom Plate of the Frame Unit by opening the eight screws.
2.
Detach the battery cables and remove the batteries.
Содержание Datex-Ohmeda S/5 Series
Страница 4: ......
Страница 90: ......
Страница 100: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 8 Document no M1144953 001 For your notes ...
Страница 118: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 26 Document no M1144953 001 For your notes ...
Страница 124: ......
Страница 128: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144955 001 For your notes ...
Страница 172: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 44 Document no M1144955 001 4 10 More ...
Страница 212: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 2 2 Document no M1144955 001 For your notes ...
Страница 214: ......
Страница 218: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144956 003 ...
Страница 266: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 48 Document no M1144956 003 ...
Страница 272: ......
Страница 304: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 4 4 Document no M1144958 001 For your notes ...
Страница 306: ......
Страница 308: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors ii Document no M1144959 001 ...