Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
24
Document no
.
M1144953-001
2.2.22 Network Option
1.
Check that the Mon-Net cable connector and the Identification plug are clean and intact,
then connect them to the frame unit. Check that the monitor connects to the
Datex-Ohmeda Network, i.e. the network symbol appears on the upper right-hand corner
of the screen. Also a message regarding the connected Central should appear in the
message field on the screen.
NOTE: If necessary, reselect the monitor’s network communication according to the used
network software in the
Network
service menu.
2.
Enter the
Network Ethernet
service menu:
Check that the counters for data errors (CRC, Frame, Transm.) are stable.
Check that the counters for hardware errors (Intern., Missed, FIFO, Overrun) all show 0.
"
2.2.23 Wireless Network Option
1.
Check that the green LED on the wireless antenna card blinks and the WLAN signal
strenght symbol scrolls between zero and full or stays fixed on the monitor screen.
2.
Check that the wireless LAN network symbol appears on the upper right-hand corner of
the screen when the monitor connects to the Datex-Ohmeda Network,
NOTE: If the monitor does not connect to the Datex-Ohmeda Network, check the WLAN
configuration on the monitor and on the network.
"
2.2.24 Interface Module, E-INT, M-INT
1.
Enter the
Interface
service menu:
Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing
faster than by 5 per second. Check that the memories have passed the internal memory
test, i.e. RAM and ROM state OK.
"
2.2.25 Device Interfacing Solution, N-DISxxx
1.
Enter the
DIS Interfacing
service menu:
Check that the DIS module ‘tout’ and ‘cse’ values do not increase faster than by 5 per
second. Check also that the DIS module memories have passed the internal memory test,
i.e. Ram, Rom and EEPROM state all OK.
Perform the same check for all connected DIS modules.
"
Содержание Datex-Ohmeda S/5 Series
Страница 4: ......
Страница 90: ......
Страница 100: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 8 Document no M1144953 001 For your notes ...
Страница 118: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 26 Document no M1144953 001 For your notes ...
Страница 124: ......
Страница 128: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144955 001 For your notes ...
Страница 172: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 44 Document no M1144955 001 4 10 More ...
Страница 212: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 2 2 Document no M1144955 001 For your notes ...
Страница 214: ......
Страница 218: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144956 003 ...
Страница 266: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 48 Document no M1144956 003 ...
Страница 272: ......
Страница 304: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 4 4 Document no M1144958 001 For your notes ...
Страница 306: ......
Страница 308: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors ii Document no M1144959 001 ...