![GE Datex-Ohmeda S/5 Series Скачать руководство пользователя страница 112](http://html.mh-extra.com/html/ge/datex-ohmeda-s-5-series/datex-ohmeda-s-5-series_technical-reference-manual_128583112.webp)
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
20
Document no
.
M1144953-001
2.2.13 Pressure/Pressure Temp Modules, E-P, E-PT, M-P, M-PT
1.
Enter the
P/PT
module service menu:
Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing
faster than by 5 per second. Check that the PT board memories have passed the internal
memory test, i.e. RAM, ROM and EEPROM all show OK.
"
Invasive blood pressure measurement
2.
Check the function of the front panel
Zero P3
membrane key.
3.
Check the InvBP channel with a patient simulator.
Zero the InvBP channel. Then check that the values and waveforms correspond to the
simulator settings.
"
Modules w/ temperature measurement
4.
Check the temperature measurement calibration using temperature test plugs.
NOTE: Make sure that the protection for temperature calibration is set on.
"
2.2.14 Dual pressure Module, E-PP, M-PP
1.
Enter the
PP
module service menu:
Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing
faster than by 5 per second. Check that the Dual Pressure module, E-PP or M-PP,
memories have passed the internal memory test, i.e. RAM, ROM and EEPROM all show OK.
2.
Check the function of the front panel
Zero P5
and
Zero P6
membrane keys.
3.
Check the InvBP channels with a patient simulator.
Zero the InvBP channels. Then check that the values and waveforms correspond to the
simulator settings.
"
2.2.15 Cardiac Output Modules, E-COP, E-COPSv, M-COP, M-COPSv
1.
Enter the
COP
module service menu:
Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing
faster than by 5 per second. Check that the module’s memories have passed the internal
memory test, i.e. RAM, ROM and EEPROM all show OK.
"
Invasive blood pressure measurement
2.
Check the function of the front panel membrane
Zero P4
key.
3.
Check the InvBP channel with a patient simulator.
Содержание Datex-Ohmeda S/5 Series
Страница 4: ......
Страница 90: ......
Страница 100: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 8 Document no M1144953 001 For your notes ...
Страница 118: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 26 Document no M1144953 001 For your notes ...
Страница 124: ......
Страница 128: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144955 001 For your notes ...
Страница 172: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 44 Document no M1144955 001 4 10 More ...
Страница 212: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 2 2 Document no M1144955 001 For your notes ...
Страница 214: ......
Страница 218: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144956 003 ...
Страница 266: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 48 Document no M1144956 003 ...
Страница 272: ......
Страница 304: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 4 4 Document no M1144958 001 For your notes ...
Страница 306: ......
Страница 308: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors ii Document no M1144959 001 ...