Aisys
2-56
04/08 M1046983
2.12.4 Fresh gas and O
2
flush flow (with SCGO)
To ABS (Circle)
breathing system
Fresh gas (
1
) flows from the electronic vaporizer (
EV
) to the SCGO assembly.
With the Circle system selected, fresh gas flow is channeled to Port 3 of the breathing
system (before the inspiratory check valve).
The output of the O
2
Flush regulator (
2
) is channeled to the O
2
Flush valve.
When activate, O
2
flush flow joins the fresh gas flow in the SCGO assembly.
Figure 2-43 • Fresh gas and O
2
flush flow (to ABS)
1
ABS
SCGO
Assembly
2
AB.75.030
EV
Circle or CGO
Selector Switch
(SCGO)
150 cm H2O
FG Pressure
Limiting Valve
SCGO
Variant
Pilot
3
2
1
O2
Sensor
5.4 psi
O2 Flush
Switch
AB.91.118
Insp Flow
Paw Trans
From
O
2
Supply
From
Mixer
To
Mixer
To
Scavenging
Absorber
2
1
Содержание Datex-Ohmeda Aisys Carestation
Страница 1: ...Aisys Anesthesia Machine Technical Reference Manual ...
Страница 16: ...Notes xiv 04 08 M1046983 ...
Страница 32: ...Notes 1 16 04 08 M1046983 ...
Страница 35: ...2 Theory of Operation M1046983 04 08 2 3 A M AB 75 098 L H D AB 75 046 J G I F K P O E D B C H N ...
Страница 158: ...Notes 5 20 04 08 M1046983 ...
Страница 288: ...Notes 7 106 04 08 M1046983 ...
Страница 325: ...9 Repair Procedures M1046983 04 08 9 29 The rear inverter A B C ...
Страница 356: ...Notes 9 60 04 08 M1046983 ...
Страница 397: ...10 Illustrated Parts M1046983 04 08 10 41 1 3 4 2 17 15 16 5 6 13 non DES cassette DES cassette 7 8 11 12 10 5 14 18 9 ...
Страница 399: ...10 Illustrated Parts M1046983 04 08 10 43 14 1 2 3 6 5 4 5 8 7 15 16 15 12 17 10 9 11 13 or 18 ...
Страница 401: ...10 Illustrated Parts M1046983 04 08 10 45 14 1 2 3 6 5 4 5 8 7 15 16 18 17 10 9 11 12 13 or ...
Страница 403: ...10 Illustrated Parts M1046983 04 08 10 47 14 15 18 1 2 3 6 5 4 5 8 7 17 10 9 11 15 16 12 13 or 17a 17c 17d 17b 17e ...
Страница 452: ...Notes 10 96 04 08 M1046983 ...
Страница 466: ...11 14 04 08 M1046983 Notes ...
Страница 470: ...Aisys 12 4 04 08 M1046983 12 3 System Schematics 12 3 1 Power Schematic ...
Страница 478: ...Aisys 12 12 04 08 M1046983 12 6 3 Tools Transfer Logs ...
Страница 495: ...12 Service Application M1046983 04 08 12 29 12 10 2 Vaporizer Subsystem Perform Vaporizer Test ...
Страница 497: ......