Aisys
5-14
04/08 M1046983
5.4.3 Inspiratory Flow Valve Cal
Calibration setup
Leave the Drive Port (
A
) and the Manifold Port (
B
) on the interface valve
plugged.
Calibration procedure:
1. On the Calibration menu, select
Insp Flow Valve
.
2. Push the ComWheel to enable the
Stage 1
calibration.
3. When Stage 1 is completed, remove the plug from the Manifold port and
insert the calibrated orifice (
C
)
4. Push the ComWheel to enable the
Stage 2
calibration.
(May take two minutes before you see any effects of the test on the screen.)
5. When Stage 2 is completed, select
Previous Menu
.
Troubleshooting
Stage 1 Calibration Failures
The Calibration will fail if the:
• Flow valve DAC counts are ≥ 1000 counts while finding points 2 through 6.
• Previously found DAC value is ≥ to the current DAC value while finding points
2 through 6.
• Previously recorded flow for a previous DAC is > the previous flow for a
previous DAC.
• Points 2, 3, and 4 have the same value stored for flow (this would cause a
divide by zero when extrapolating).
Possible causes for calibration failure:
• Check Altitude and Drive Gas selection
• Leaks around the test plugs
• Leaks in Vent Engine Interface Manifold – Inspect for leaks
• Insp Flow Valve not closing completely (leaky) – Replace Insp Flow Valve
• Drive gas regulator not adjusted / stable – Check regulator calibration
• Insp Flow Valve not linear – Replace Insp Flow Valve
A
B
A
C
C
Calibrated Flow Orifice
1504-3016-000
Содержание Datex-Ohmeda Aisys Carestation
Страница 1: ...Aisys Anesthesia Machine Technical Reference Manual ...
Страница 16: ...Notes xiv 04 08 M1046983 ...
Страница 32: ...Notes 1 16 04 08 M1046983 ...
Страница 35: ...2 Theory of Operation M1046983 04 08 2 3 A M AB 75 098 L H D AB 75 046 J G I F K P O E D B C H N ...
Страница 158: ...Notes 5 20 04 08 M1046983 ...
Страница 288: ...Notes 7 106 04 08 M1046983 ...
Страница 325: ...9 Repair Procedures M1046983 04 08 9 29 The rear inverter A B C ...
Страница 356: ...Notes 9 60 04 08 M1046983 ...
Страница 397: ...10 Illustrated Parts M1046983 04 08 10 41 1 3 4 2 17 15 16 5 6 13 non DES cassette DES cassette 7 8 11 12 10 5 14 18 9 ...
Страница 399: ...10 Illustrated Parts M1046983 04 08 10 43 14 1 2 3 6 5 4 5 8 7 15 16 15 12 17 10 9 11 13 or 18 ...
Страница 401: ...10 Illustrated Parts M1046983 04 08 10 45 14 1 2 3 6 5 4 5 8 7 15 16 18 17 10 9 11 12 13 or ...
Страница 403: ...10 Illustrated Parts M1046983 04 08 10 47 14 15 18 1 2 3 6 5 4 5 8 7 17 10 9 11 15 16 12 13 or 17a 17c 17d 17b 17e ...
Страница 452: ...Notes 10 96 04 08 M1046983 ...
Страница 466: ...11 14 04 08 M1046983 Notes ...
Страница 470: ...Aisys 12 4 04 08 M1046983 12 3 System Schematics 12 3 1 Power Schematic ...
Страница 478: ...Aisys 12 12 04 08 M1046983 12 6 3 Tools Transfer Logs ...
Страница 495: ...12 Service Application M1046983 04 08 12 29 12 10 2 Vaporizer Subsystem Perform Vaporizer Test ...
Страница 497: ......