Revision A
Dash 3000/4000 Patient Monitor
9-51
2000966-339
Assembly Drawings:
Port Connections
Network Interface Connector
An 8-pin RJ-45 connector containing two isolated, differential pairs is
provided to connect the monitor to a network hub.
Auxiliary Communication Connector
Auxiliary communication communicates with peripherals such as a
remote control or data logger. The UART channel is only capable of
communicating with external devices having the same transmit and
receive baud rates. Power available to peripheral devices is current
limited to 100 mA.
PIN
SIGNAL NAME
I/O
SIGNAL DESCRIPTION
DIAGRAM
1
LAN _TX+
O
LAN tr
2
LAN_TX-
O
LAN transmit –
3
LAN _RX+
O
LAN r
4
NC
–
No connection
5
NC
–
No connection
6
LAN _RX-
O
LAN receive –
7
NC
–
No connection
8
NC
–
No connection
8 7 6 5 4 3 2 1
585A
PIN
SIGNAL NAME
I/O
SIGNAL DESCRIPTION
DIAGRAM
1
AU5V
O
Is5V
2
SPARE
I
Spare
3
AUTOPORT_232_RX
I
RS-232 receive signal
4
AUTOPORT_GND
–
Power return
5
AUTOPORT_ID
I/O
Device identification signal - host side
6
AUTOPORT_232_TX
O
RS-232 transmit signal
7
AU10V_RTS
O
RS-232 request-to-send signal (remote alarm
control signal)
8
AU10V_DSR
O
RS-232 data-terminal-ready signal
8 7 6 5 4 3 2 1
585A
Содержание Dash 3000
Страница 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Страница 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Страница 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Страница 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Страница 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Страница 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Страница 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Страница 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Страница 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 236: ...9 12 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Theory Of Operation Block Diagram ...
Страница 278: ...9 54 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Port Connections For your notes ...
Страница 279: ...Revision A Dash 3000 4000 Patient Monitor A 1 2000966 339 A Appendix A Electromagnetic Compatibility EMC ...
Страница 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 291: ......