Managing Patients
Discharge
Discharging a patient from a monitor clears all patient information while retaining the
unit name, bed number and factory or custom defaults.
WARNING
DISCHARGE TO CLEAR PATIENT DATA—When admitting a
new patient to the monitor, you must clear all previous patient data
from the monitor. To accomplish this, disconnect all patient cables
and discharge the patient from the monitor.
To manually discharge a patient from the monitor, follow this procedure:
1.
Disconnect all patient cables.
2.
Select
MORE MENUS
>
ADMIT MENU
>
DISCHARGE PATIENT
or
NEW
CASE
.
3.
If in
COMBO
or
ROVER
COMBO
monitoring modes, select on of the following
to terminate telemetry monitoring:
MONITOR
— Discharge from monitor only. If ECG was acquired from
telemetry, telemetry will be monitored by the central station.
TELEMETRY
— Discharge from telemetry only. ECG monitoring
automatically switches to the monitor.
BOTH
— Discharge from monitor and telemetry.
4.
If the monitor is setup to view other monitors, the following events occur when
the host monitor is discharged:
When running the VOA or AVOA software option, the viewed bed is
removed from the display and the monitor reverts to the factory or custom
defaults.
When running the AutoView XM software option, the viewed bed is
removed from the display, the monitor reverts to the factory or custom
defaults. All monitors from other care units are removed from the list of
watched monitors. All monitors in the same care unit are retained and
monitored according to the
VIEW ON ALARM
control settings.
SET LOCATION ID
ROVER
and
ROVER
COMBO
All
Select the MUSE Cardiology Information System for network
identification.
ECG SOURCE
COMBO
and
ROVER
COMBO
All
Select the monitor or telemetry transmitter to receive ECG data. If
you select telemetry you must also select the transmitter number
before you can admit a patient to the monitor.
1
All includes all four monitoring modes; STANDARD, ROVER, COMBO and ROVER COMBO.
2
All includes all three patient-monitor types; ADULT-ICU, NEONATAL-ICU and OPERATING ROOM.
Option
Monitoring
Mode
Patient-Monitor
Type
Function
Содержание Dash 3000
Страница 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Страница 12: ...x Dash 3000 4000 5000 2000966 386D ...
Страница 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Страница 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Страница 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Страница 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Страница 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Страница 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Страница 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Страница 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Страница 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Страница 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Страница 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Страница 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Страница 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Страница 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Страница 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Страница 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Страница 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Страница 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Страница 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Страница 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Страница 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Страница 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Страница 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Страница 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Страница 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Страница 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Страница 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Страница 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Страница 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Страница 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Страница 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Страница 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Страница 269: ......