Measuring Cardiac Output
Performing the CO Procedure
To perform a CO procedure, follow this procedure. Before you begin this procedure,
verify you are ready by reviewing
1.
Select the
CO
parameter window label.
2.
If the monitor is configured for automatic measurements, the “
INJECT WHEN
READY
” message and
CARDIAC OUTPUT
information window displays.
3.
Watch for the
“Inject Now
” message and select
CARDIAC OUTPUT NOW
to
calculate the cardiac output.
The information window displays the current cardiac trial information, a review
CO curve, the derived CO value, the average CO value between stored trials, the
patient’s heart rate and other vital signs at the time of the trial. It also includes
cardiac disturbances (
RHY
), the patient’s height and weight and calculated
parameters.
PRINT CO CURVE
Print the displayed real-time waveform. A “
PRINTING
MANUAL
” message displays when the waveform is printing.
AUTO MODE
Turn automatic measurements on or off. The monitor
searches for a stable baseline by averaging approximately
8.5 seconds of the patient’s blood temperature and displays
the message “
INJECT WHEN READY
” when a stable
baseline is established.
USE
Before the injection procedure begins, select one of the
blood pressure values to be factored in the cardiac
calculation:
CATHETER
Before the injection procedure begins, select the correct
catheter type. The monitor guides you through the
INJECT
TEMP
,
SIZE
and
INJECT VOL
options to assure correct
setup.
If you select
OTHER
, you must select a value for
COMPUTATION CONSTANT
. Refer to the literature
provided by the catheter manufacturer for the correct value.
INJECT TEMP
Select the injectate probe type.
SIZE
Select the catheter French size.
INJECT VOL
Select the injectate volume.
COMPUTATION
CONSTANT
Select a computation constant for a catheter manufacturer.
If you selected a specific manufacturer for the
CATHETER
,
the correct value for the manufacturer is automatically
selected for you.
If you selected
OTHER
for the
CATHETER
, you must select
a value or you will not be able to complete a trial. Refer to
the literature provided by the catheter manufacturer for the
correct value.
CARDIAC OUTPUT HELP
Display troubleshooting information.
Option
Function
Содержание Dash 3000
Страница 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Страница 12: ...x Dash 3000 4000 5000 2000966 386D ...
Страница 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Страница 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Страница 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Страница 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Страница 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Страница 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Страница 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Страница 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Страница 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Страница 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Страница 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Страница 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Страница 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Страница 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Страница 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Страница 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Страница 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Страница 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Страница 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Страница 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Страница 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Страница 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Страница 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Страница 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Страница 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Страница 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Страница 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Страница 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Страница 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Страница 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Страница 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Страница 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Страница 269: ......