4-52
250cx Series Maternal/Fetal Monitor
Revision C
2036947-001
Maintenance: Recorder Photosensor Calibration
Recorder Photosensor Calibration
Adjusting the Paper-Low Photosensor
1.
Load paper in the recorder. Ensure that there are
no
black squares showing to
indicate a paper-low condition.
2.
Press the
Record
button to turn on the recorder. Allow the paper to advance for
a few seconds in order to tension the paper.
3.
Turn off the recorder.
4.
Using a digital voltmeter, connect the positive lead to J9 (pin 4) on the Recorder
Board; connect the negative lead to J9 (pin 2).
5.
Adjust R31 for a reading of +150 mVdc ± 2 mVdc on the digital voltmeter.
NOTE:
If you open and then close the recorder door, the reading may vary 5–10
mV, due to the loss of tension in the paper. This is acceptable and you do
not need to re-adjust.
6.
Open the door and verify that the reading on the digital voltmeter is greater than
+4.75 Vdc.
7.
Re-load the paper so that black squares show on the surface (i.e., the last several
sheets of a pack).
8.
Turn on the recorder.
9.
The value on the digital voltmeter should go up and down as the paper surface
alternates between black and white. Verify that the maximum value is greater
than or equal to 2.0 Vdc.
10. Turn off the recorder.
Adjusting the Paper-Out Photosensor
1.
Load paper in the recorder. Ensure that there are
no
black squares showing to
indicate a paper-low condition.
2.
Press the
Record
button to turn on the recorder. Allow the paper to advance for
a few seconds in order to tension the paper.
3.
Turn off the recorder.
4.
Using a digital voltmeter, connect the positive lead to J9 (pin 3) on the Recorder
Board; connect the negative lead to J9 (pin 2).
5.
Adjust R29 for a reading of +150 mVdc ± 2 mVdc on the digital voltmeter.
NOTE:
If you open and then close the recorder door, the reading may vary 5–10
mV, due to the loss of tension in the paper. This is acceptable and you do
not need to re-adjust.
6.
Open the door and verify that the reading on the digital voltmeter is greater than
+4.75 Vdc.
7.
Turn off the recorder.
Содержание Corometrics 250cx Series
Страница 2: ......
Страница 6: ...CE CE ii 0086 ...
Страница 14: ...viii 250cx Series Maternal Fetal Monitor Revision C 2036947 001 ...
Страница 16: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Страница 28: ...1 14 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Introduction Service Requirements ...
Страница 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036947 001 2 Equipment Overview ...
Страница 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Страница 54: ...2 26 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Equipment Overview Theory of Operation ...
Страница 55: ...3 1 250cx Series Maternal Fetal Monitor Revision C 2036947 001 3 Installation ...
Страница 56: ...3 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Страница 81: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036947 001 4 Maintenance ...
Страница 82: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Страница 140: ...4 60 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Maintenance Preventative Maintenance Inspection Report ...
Страница 141: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036947 001 5 Troubleshooting ...
Страница 142: ...5 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Страница 192: ...5 52 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Troubleshooting General Troubleshooting ...
Страница 193: ...Revision C 250cx Series Maternal Fetal Monitor 6 1 2036947 001 6 Parts List Drawings and Replacement ...
Страница 194: ...6 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Страница 226: ...A 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Technical Specifications For your notes ...
Страница 236: ...A 12 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Technical Specifications Strip Chart Recorder ...
Страница 237: ...Revision C 250cx Series Maternal Fetal Monitor B 1 2036947 001 B Alarms Summary ...
Страница 238: ...B 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Alarms Summary For your notes ...
Страница 241: ...Revision C 250cx Series Maternal Fetal Monitor C 1 2036947 001 C Electromagnetic Compatibility ...
Страница 242: ...C 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Electromagnetic Compatibility For your notes ...
Страница 250: ...C 10 250cx Series Maternal Fetal Monitor Revision C 2036947 001 This page intentionally left blank ...
Страница 251: ......