5-16
170 Series Monitor
Revision C
2000947-004
Theory of Operation:
Functional Overview
* Active low.
Table 5-17. FECG/UA Option Interface Part Two (J15), Models 173/174 Only
Pin Number
Signal Name
Signal Type
(Relative To Main
Board)
Signal Description
1
RA
Output
Right Arm Input from Patient Connector
2
LA
Output
Left Arm Input from Patient Connector
3
RL
Input
Right Leg Drive Signal from FECG/IUP Board
4
SHIELDF
Output
Isolated Shield
5
–PRESS
Output
UA Channel Negative Input
6
+PRESS
Output
UA Channel Positive Input
7
ISO4V
Input
Isolated 4 V Reference
8
ECGEN*
Output
Enable for FECG Cable
9
4VISOGND
Input
Isolated 4 V Reference Ground
10
PEN*
Output
Isolated IUP Enable
11
NC
—
No Connection
12
ENTOCO*
Output
Isolated TOCO Enable
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 18: ...1 10 170 Series Monitor Revision C 2000947 004 Safety Equipment Symbols For your notes ...
Страница 24: ...2 6 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 40: ...3 16 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 59: ...Revision C 170 Series Monitor 4 19 2000947 004 Setup Procedures Quick Reference Card ...
Страница 112: ...6 10 170 Series Monitor Revision C 2000947 004 Functional Checkout Procedure FECG Test Figure 6 1 FECG Ramp ...
Страница 117: ...Revision C 170 Series Monitor 6 15 2000947 004 Functional Checkout Procedure Ultrasound Test Figure 6 3 Ultrasound Ramp ...
Страница 182: ...7 50 170 Series Monitor Revision C 2000947 004 Serviceable Assemblies Boot ROM Error Codes For your notes ...
Страница 202: ...8 20 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 224: ...9 22 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 247: ......