4-8
170 Series Monitor
Revision C
2000947-004
Setup Procedures:
Monitor Self-Test Routines
Monitor Self-Test Routines
NOTE:
Ensure paper is installed in the recorder in order to verify a successful
recorder test.
Each 170 Series Monitor contains a self-test routine which checks the internal
circuitry of the monitor, the displays and indicators, and the strip chart recorder.
The self-test routine is automatically initiated each time you turn on the monitor.
CAUTION
SELF-TEST FAILURE—If there is any problem with the self-test
routine, turn off the monitor and remove it from operation. Notify
your Biomedical Engineering Department or Service
Representative.
After completion of a successful self-test routine, the monitor is ready for use.
NOTE:
If the recorder was off at the time the monitor was turned off, the test
routine will turn the recorder on, then turn it off after the tests are complete.
If the recorder was on at the time the monitor was turned off, the tests will
be performed and the recorder will remain on.
Table 4-1. Summary of Self-Test Routines
Test Description
What to Verify
Display/Indicator Test:
All displays and indicators
illuminate.
Ensure all indicators and each segment of the displays
illuminate throughout the entire self-test routine.
Internal Test:
The internal circuitry of the monitor is
verified.
Make sure the monitor performs the recorder test. If
there is a problem with the internal circuitry, the recorder
test will not be performed.
Recorder Test:
The following message prints on the
strip chart paper: TEST: ARE ALL DOTS PRINTED?
Three continuous lines are drawn across the strip chart
recorder paper, testing the integrity of the printhead.
See
.
Ensure that the lines are printed in the correct positions
on the paper. Verify that the lines are continuous and no
gaps appear on the traces.
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 18: ...1 10 170 Series Monitor Revision C 2000947 004 Safety Equipment Symbols For your notes ...
Страница 24: ...2 6 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 40: ...3 16 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 59: ...Revision C 170 Series Monitor 4 19 2000947 004 Setup Procedures Quick Reference Card ...
Страница 112: ...6 10 170 Series Monitor Revision C 2000947 004 Functional Checkout Procedure FECG Test Figure 6 1 FECG Ramp ...
Страница 117: ...Revision C 170 Series Monitor 6 15 2000947 004 Functional Checkout Procedure Ultrasound Test Figure 6 3 Ultrasound Ramp ...
Страница 182: ...7 50 170 Series Monitor Revision C 2000947 004 Serviceable Assemblies Boot ROM Error Codes For your notes ...
Страница 202: ...8 20 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 224: ...9 22 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 247: ......