7-20
170 Series Monitor
Revision C
2000947-004
Serviceable Assemblies:
Tocotransducer Calibration
Tocotransducer Calibration
Any of the following three transducers can be used with the 170 Series Monitor:
Corometrics Nautilus Tocotransducer
Corometrics Trimline Tocotransducer
Corometrics Large-Button Tocotransducer
Figure 7-7. CMR Test Jack
Table 7-8. CMR Test Jack Components
Part Description
Catalog Number
Pressure Connector
6158J
475
Ω
, 1/8 W, 1% Resistor
472127
+P
P
+4 V
REF
Pin 1*
Pin 2*
Pin 4*
Pin 6*
475
W
*Front Panel UA Connector
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 18: ...1 10 170 Series Monitor Revision C 2000947 004 Safety Equipment Symbols For your notes ...
Страница 24: ...2 6 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 40: ...3 16 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 59: ...Revision C 170 Series Monitor 4 19 2000947 004 Setup Procedures Quick Reference Card ...
Страница 112: ...6 10 170 Series Monitor Revision C 2000947 004 Functional Checkout Procedure FECG Test Figure 6 1 FECG Ramp ...
Страница 117: ...Revision C 170 Series Monitor 6 15 2000947 004 Functional Checkout Procedure Ultrasound Test Figure 6 3 Ultrasound Ramp ...
Страница 182: ...7 50 170 Series Monitor Revision C 2000947 004 Serviceable Assemblies Boot ROM Error Codes For your notes ...
Страница 202: ...8 20 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 224: ...9 22 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 247: ......