![GE Corometrics 170 Series Скачать руководство пользователя страница 138](http://html.mh-extra.com/html/ge/corometrics-170-series/corometrics-170-series_service-manual_128350138.webp)
7-6
170 Series Monitor
Revision C
2000947-004
Serviceable Assemblies:
Transducer Plug Replacement Kits
Figure 7-1. Tocotransducer Cable Preparation
1
²
²
18
¤
1
²
12
¤
1
²
yellow
#24 AWG Green Jumper Wire
#24 AWG Green Wire
²
18
¤
Oetiker Clamp
(Nautilus only)
Vent Tube (Nautilus only)
Shrink Tubing
Shield
Table 7-3. Tocotransducer Cable Wiring
Pin Number
Pin Length
Description
Wire Color
1
short
+PRESS
Red
2
short
–PRESS
White
3
short
—
—
4
short
+4 V
Blue
5 short
—
—
6
short
Reference/Enable
#24 AWG Green Jumper
and
Yellow
7
LONG
Shield Ground
#24 AWG Green
8
short
—
—
9
short
—
—
10
short
—
—
11
short
—
—
12
short
TOCO Enable
#24 AWG Green Jumper
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 18: ...1 10 170 Series Monitor Revision C 2000947 004 Safety Equipment Symbols For your notes ...
Страница 24: ...2 6 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 40: ...3 16 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 59: ...Revision C 170 Series Monitor 4 19 2000947 004 Setup Procedures Quick Reference Card ...
Страница 112: ...6 10 170 Series Monitor Revision C 2000947 004 Functional Checkout Procedure FECG Test Figure 6 1 FECG Ramp ...
Страница 117: ...Revision C 170 Series Monitor 6 15 2000947 004 Functional Checkout Procedure Ultrasound Test Figure 6 3 Ultrasound Ramp ...
Страница 182: ...7 50 170 Series Monitor Revision C 2000947 004 Serviceable Assemblies Boot ROM Error Codes For your notes ...
Страница 202: ...8 20 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 224: ...9 22 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 247: ......