6-6
CIC Pro™
2026419-033E
Configuration
WARNING
During the timeframe that GE patient monitoring devices are
connected to a non-validated network, customers must be aware that
they are operating this system with increased risk, especially for
devices that rely specifically on the network for real time
transmission of alarms and other monitoring data. This increased
risk comes about because GE has not been able to review/approve
the proposed network design and/or commission the implemented
network to ensure it meets required performance specifications.
The devices at increased risk include, but are not exclusive to,
medical telemetry and all patient monitoring done from the CIC Pro
center.
Since monitoring data flows to and from central stations, telemetry
servers, and other medical devices, the lack of a commissioned
network can affect the performance of the overall system. Further,
the lack of a validated network may result in limited technical
support for troubleshooting product issues on products that rely on
the CARESCAPE Network.
8.
Investigate if there are any spare (un-connected) CIC Pro centers in the biomed
shop and/or if there are any CIC Pro centers in the care area that are designed to
work on customer demand (e.g., connected to the network, currently switched off
and used only when needed). Before introducing such CIC Pro centers to the
existing network, do the following:
a.
Make sure that those CIC Pro center hardware and software versions are
compatible (as listed in step
The time zone settings are the same.
The
Automatically adjust clock for daylight saving changes
check
box must remain
UNCHECKED
at all times.
All MC IP addresses must be in the same IP scheme with the same
subnet mask.
All IX IP addresses must be in the same IP scheme with the same subnet
mask.
b.
Inform the biomed about the following Caution:
CAUTION
NETWORK DEVICE TIME SYNCHRONIZATION — When
adding a new device (e.g., CIC Pro center) to the CARESCAPE
Network, the existing devices on the CARESCAPE Network will
synchronize to the new device’s time. To prevent potential time
synchronization issues, you should set the new device’s time to be as
close as possible to the time (within one minute or less) used by the
existing GE devices on the CARESCAPE Network.
9.
Make CIC v5.1.x (or the highest CIC Pro center software version) the Time
Master.
Содержание CIC Pro
Страница 12: ...x CIC Pro 2026419 033E ...
Страница 13: ...2026419 033E CIC Pro 1 1 1 Introduction ...
Страница 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview ...
Страница 53: ...2026419 033E CIC Pro 3 1 3 Licensing ...
Страница 60: ...3 8 CIC Pro 2026419 033E Licensing ...
Страница 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces ...
Страница 74: ...4 14 CIC Pro 2026419 033E Service interfaces ...
Страница 75: ...2026419 033E CIC Pro 5 1 5 Installation ...
Страница 94: ...5 20 CIC Pro 2026419 033E Installation ...
Страница 95: ...2026419 033E CIC Pro 6 1 6 Configuration ...
Страница 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures ...
Страница 208: ...7 22 CIC Pro 2026419 033E Checkout procedures ...
Страница 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting ...
Страница 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs ...
Страница 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive ...
Страница 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount ...
Страница 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options ...
Страница 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance ...
Страница 279: ...2026419 033E CIC Pro 11 1 11 Reload software ...
Страница 288: ...11 10 CIC Pro 2026419 033E Reload software ...
Страница 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software ...
Страница 298: ...12 10 CIC Pro 2026419 033E Upgrade software ...
Страница 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility ...
Страница 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules ...
Страница 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility ...
Страница 312: ...C 4 CIC Pro 2026419 033E Device compatibility ...
Страница 313: ......