Setting MV/TV alarms
Use the
MV/TV Alarms
setting to turn off the MV and TV alarms.
Settings made during manual ventilation are not retained when
mechanical ventilation starts. Settings made during mechanical
ventilation are retained when manual ventilation starts.
For example, if
MV/TV Alarms
is set to
Off
during manual
ventilation, the alarms remain off until the Bag/Vent switch is set to
Vent or the
MV/TV Alarms
is set to
On
.
For example, if the
MV/TV Alarms
is set to
Off
during mechanical
ventilation, the alarms remain off when manual ventilation starts.
1.
Select
Alarm Setup
.
2.
To turn the volume alarms off, set
MV/TV Alarms
to
Off
.
‘MV/TV Alarms Off’ appears in the general message field.
The volume alarm limits waveform numerics shows alarm off
symbol during a case.
3.
To turn the volume alarms on, set
MV/TV Alarms
to
On
.
4.
Select
Close
.
Setting alarm limits
Do not set alarm limits to extreme values. Setting limits to
extreme values can render the alarm useless.
1.
Select
Alarm Setup
.
2.
From the
Primary Limits
and
More Limits
tabs, select the
alarm limit and make the change.
3.
Push the Home key, touch the waveform area of the display, or
select
Close
to close the menu.
Viewing alarm history
Use the
Alarm History
tab to view the list of the 12 most recent high
and medium priority alarms that occurred since the start of the case.
The alarm history clears at the start of a new case.
1.
Select
Alarm Setup
.
2.
Select the
Alarm History
tab.
The list of alarms shows in the window.
WARNING
Carestation
™
620/650/650c (A1)
3-22
2076152-001
Содержание Carestation 620
Страница 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Страница 12: ...Carestation 620 650 650c A1 x 2076152 001...
Страница 57: ...3 Push the ComWheel or select the quick key to confirm the change 2 System controls and menus 2076152 001 2 23...
Страница 58: ...2 24 2076152 001...
Страница 98: ...3 40 2076152 001...
Страница 102: ...4 4 2076152 001...
Страница 128: ...6 12 2076152 001...
Страница 150: ...7 22 2076152 001...
Страница 178: ...4 Tighten all thumb screws on the rear of the mounting plate AC 24 265 Carestation 620 650 650c A1 8 28 2076152 001...
Страница 182: ...8 32 2076152 001...
Страница 190: ...12 Calibrate the new O2 cell according to the Circuit O2 cell procedure Carestation 620 650 650c A1 9 8 2076152 001...
Страница 196: ...9 14 2076152 001...
Страница 246: ...11 44 2076152 001...
Страница 250: ...3 Set the cost for each agent 4 Select Back to return to the Gas Usage menu Carestation 620 650 650c A1 12 4 2076152 001...
Страница 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...