Setting CO2 alarms
Use the CO2 Alarms setting to turn off the CO2 ‘Apnea’ alarm,
‘EtCO2 low’, ‘EtCO2 high’, and ‘FiCO2 high. Absorbent OK?’ alarms
during manual ventilation.
CO2 information is obtained from the airway module installed in the
anesthesia system module bay. The CO2 alarms setting has no
effect if there is no airway module in the anesthesia system.
1.
Select
Alarm Setup
.
2.
To turn off the CO2 alarms, set
CO2 Alarms
to
Off
.
The ‘CO2 Alarms Off’ message shows in the general message
field.
The CO2 alarm limit waveform numerics shows alarm off symbol
during a case.
The alarms remain disabled until the Bag/Vent switch is set to
Vent, the case is ended, or the
CO2 Alarms
is set to
On
.
3.
To turn on the CO2 alarms, set
CO2 Alarms
to
On
.
4.
Select
Close
.
Setting volume apnea alarm
Use the
Vol Apnea Alarm
setting to turn off the volume apnea alarm
during manual ventilation. The volume apnea alarm remains off until
the Bag/Vent switch is set to Vent or
Vol Apnea Alarm
is set to
On
.
Vol Apnea Alarm
does not show on the
Alarm Setup
menu when
the
Volume Apnea Selection
has been disabled by the Super User.
" section for more information.
1.
Select
Alarm Setup
.
2.
To turn the volume apnea alarms off, select
Vol Apnea Alarm
to
Off
.
‘Volume Apnea Off’ shows in the general message field.
•
If mechanical ventilation is started, the volume apnea
alarms are active.
•
If manual ventilation is restarted, a pop-up confirmation
window appears to resume the
Off
setting.
3.
To turn the volume apnea alarms on, set
Vol Apnea Alarm
to
On
.
4.
Select
Close
.
Note
3 Operation
2076152-001
3-21
Содержание Carestation 620
Страница 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Страница 12: ...Carestation 620 650 650c A1 x 2076152 001...
Страница 57: ...3 Push the ComWheel or select the quick key to confirm the change 2 System controls and menus 2076152 001 2 23...
Страница 58: ...2 24 2076152 001...
Страница 98: ...3 40 2076152 001...
Страница 102: ...4 4 2076152 001...
Страница 128: ...6 12 2076152 001...
Страница 150: ...7 22 2076152 001...
Страница 178: ...4 Tighten all thumb screws on the rear of the mounting plate AC 24 265 Carestation 620 650 650c A1 8 28 2076152 001...
Страница 182: ...8 32 2076152 001...
Страница 190: ...12 Calibrate the new O2 cell according to the Circuit O2 cell procedure Carestation 620 650 650c A1 9 8 2076152 001...
Страница 196: ...9 14 2076152 001...
Страница 246: ...11 44 2076152 001...
Страница 250: ...3 Set the cost for each agent 4 Select Back to return to the Gas Usage menu Carestation 620 650 650c A1 12 4 2076152 001...
Страница 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...