Breathing system specifications
Compressible volume Ventilator side
2006 ml
Bag side
500 ml
Reusable canister
1000 ml
Disposable canister
1000 ml
Absorbent
Reusable canister
1370 ml
Disposable canister
1400 ml
Connections
Auxiliary Common Gas Outlet: ISO 5356 type
connector on the front of the system (standard 22
mm OD or 15 mm ID conical friction fit connectors).
Breathing system
leakage
Less than or equal to 150 ml/min total at 3 kPa (30
cmH2O) (both in absorber mode and with canister
removed).
Breathing system
compliance
Volume of gas lost due to internal compliance of
filled reusable absorber canister (bag mode only):
1.74 ml/0.098 kPa (1 cmH2O)
52.4 ml/3 kPa (30 cmH2O)
Volume of gas lost due to internal compliance of
filled disposable absorber canister (bag mode only):
1.81 ml/0.098 kPa (1 cmH2O)
54.56 ml/3 kPa (30 cmH2O)
APL valve
Approximately 0 to 70 cmH2O
Negative pressure
relief
Minimum 100 ml/min at -14 cmH2O
Mechanical
overpressure valve
115 +/- 10 cmH2O
Inspiratory breathing resistance in bag mode*
l/min
kPa
cmH2O
Absorber canister installed
5
0.030
0.30
30
0.172
1.72
60
0.456
4.56
Absorber canister removed
5
0.034
0.34
30
0.159
1.59
60
0.401
4.01
*Values include patient circuit tubing and Y-piece 0.065 kPa (0.65 cmH2O)
at 60 l/min. Patient circuit tubing and breathing system configurations affect
resistance.
Carestation
™
620/650/650c (A1)
11-12
2076152-001
Содержание Carestation 620
Страница 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Страница 12: ...Carestation 620 650 650c A1 x 2076152 001...
Страница 57: ...3 Push the ComWheel or select the quick key to confirm the change 2 System controls and menus 2076152 001 2 23...
Страница 58: ...2 24 2076152 001...
Страница 98: ...3 40 2076152 001...
Страница 102: ...4 4 2076152 001...
Страница 128: ...6 12 2076152 001...
Страница 150: ...7 22 2076152 001...
Страница 178: ...4 Tighten all thumb screws on the rear of the mounting plate AC 24 265 Carestation 620 650 650c A1 8 28 2076152 001...
Страница 182: ...8 32 2076152 001...
Страница 190: ...12 Calibrate the new O2 cell according to the Circuit O2 cell procedure Carestation 620 650 650c A1 9 8 2076152 001...
Страница 196: ...9 14 2076152 001...
Страница 246: ...11 44 2076152 001...
Страница 250: ...3 Set the cost for each agent 4 Select Back to return to the Gas Usage menu Carestation 620 650 650c A1 12 4 2076152 001...