Cancelling audio pause
Selecting and holding
Audio Pause
for 2 seconds will cancel the
audio pause function.
Display changes during alarms
Messages may appear in the waveform field during some alarms. If
more than one alarm has a message, the message for the highest
priority alarm is displayed. The message is removed when the alarm
is resolved.
The color of the alarm text shown in the alarm message fields is
dependent on the alarm priority. Messages for high-priority alarms
use white text. Messages for medium-priority alarms use black text.
Low-priority alarm messages use black text.
When the O2 pipeline supply pressure drops to less than 252 kPa
(36 psi), O2 supply information is displayed.
Some patient parameter alarms such as ‘Ppeak high’ and ‘FiO2 low’
will latch when the alarm condition is corrected. When an alarm is
latched, it is displayed in gray text on a black background. The
parameter box will stop flashing. All the associated waveform,
numeric, and digit field messages are removed from the display. The
alarm will remain in this condition until it is acknowledged by
selecting
Audio Pause
or until the alarm reoccurs. When the alarm
is acknowledged, it is removed from the screen. If an alarm has
latched and the alarm reoccurs before it is acknowledged, the alarm
will revert to an active state.
De-escalating alarms
Some device related alarms, such as ‘No insp flow sensor’ will de-
escalate priority when the alarm is acknowledged by selecting Audio
Pause. The alarm message shows at the low-priority alarm level until
the alarm condition is resolved and the alarm is cleared. If that alarm
reoccurs after it has been resolved, the alarm occurs at its standard
priority level.
Battery indicator
The color and fill amount of the battery in use symbol indicates the
amount of battery power remaining.
•
Green indicates greater than 60 minutes.
•
Yellow indicates between 60 and 5 minutes.
7 Alarms and troubleshooting
2076152-001
7-3
Содержание Carestation 620
Страница 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Страница 12: ...Carestation 620 650 650c A1 x 2076152 001...
Страница 57: ...3 Push the ComWheel or select the quick key to confirm the change 2 System controls and menus 2076152 001 2 23...
Страница 58: ...2 24 2076152 001...
Страница 98: ...3 40 2076152 001...
Страница 102: ...4 4 2076152 001...
Страница 128: ...6 12 2076152 001...
Страница 150: ...7 22 2076152 001...
Страница 178: ...4 Tighten all thumb screws on the rear of the mounting plate AC 24 265 Carestation 620 650 650c A1 8 28 2076152 001...
Страница 182: ...8 32 2076152 001...
Страница 190: ...12 Calibrate the new O2 cell according to the Circuit O2 cell procedure Carestation 620 650 650c A1 9 8 2076152 001...
Страница 196: ...9 14 2076152 001...
Страница 246: ...11 44 2076152 001...
Страница 250: ...3 Set the cost for each agent 4 Select Back to return to the Gas Usage menu Carestation 620 650 650c A1 12 4 2076152 001...
Страница 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...