Calibrating the airway module
Calibrate airway modules once every six months or whenever there
are indications of errors in the gas readings. Use a manufacturer
approved calibration gas and regulator to calibrate the modules. See
the "
" section for the stock numbers of the calibration gas and
regulator.
Only use manufacturer approved calibration gas. Do not
use any other calibration gases or the calibration will not
succeed. Dispose of calibration gas containers in
accordance with local environmental procedures.
Calibration gas contains anesthetic agents. Make sure
that there is sufficient ventilation in the room during
airway module calibration.
Use of an airway module that has failed calibration may
cause inaccurate readings. Do not use airway modules
that fail calibration.
The
Calibration
menu is not available during
Checkout
or during a
case.
During gas calibration, % units are used for CO2
regardless of selected measuring units.
1.
Turn on the power. Let the module warm up for 30 minutes
before starting calibration.
2.
Attach the regulator to the calibration gas cylinder.
3.
Attach a new sampling line to the water trap. Connect the loose
end of the sampling line to the regulator on the calibration gas
cylinder.
4.
Select the
System Setup
button.
5.
Select
Calibration
.
6.
Select
Airway Gas
.
7.
Wait until
Connect the calibration gas and feed gas.
appears
on the anesthesia display.
•
If an error occurs during calibration or if no gas is fed, the
calibration shows ‘Fail’. Select
Back
to perform a new
calibration.
8.
For regulators with a numbered gauge, open the regulator until
the gauge reads between 5 to 7 psi (34 to 48 kPa). For
regulators with a non-numbered gauge, open the valve.
9.
Feed the calibration gas until
End gas feed. Match the values
with the calibration gas. Confirm each.
appears.
•
If an error occurs during calibration or if no gas is fed,
‘Calibration error’ appears. Push the ComWheel to perform
a new calibration.
WARNING
•
•
Note
Carestation
™
620/650/650c (A1)
6-10
2076152-001
Содержание Carestation 620
Страница 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Страница 12: ...Carestation 620 650 650c A1 x 2076152 001...
Страница 57: ...3 Push the ComWheel or select the quick key to confirm the change 2 System controls and menus 2076152 001 2 23...
Страница 58: ...2 24 2076152 001...
Страница 98: ...3 40 2076152 001...
Страница 102: ...4 4 2076152 001...
Страница 128: ...6 12 2076152 001...
Страница 150: ...7 22 2076152 001...
Страница 178: ...4 Tighten all thumb screws on the rear of the mounting plate AC 24 265 Carestation 620 650 650c A1 8 28 2076152 001...
Страница 182: ...8 32 2076152 001...
Страница 190: ...12 Calibrate the new O2 cell according to the Circuit O2 cell procedure Carestation 620 650 650c A1 9 8 2076152 001...
Страница 196: ...9 14 2076152 001...
Страница 246: ...11 44 2076152 001...
Страница 250: ...3 Set the cost for each agent 4 Select Back to return to the Gas Usage menu Carestation 620 650 650c A1 12 4 2076152 001...
Страница 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...