Connecting the airway module
1.
Check that the airway module is installed.
2.
Check that the airway adapter connections are tight and that the
adapter is correctly installed.
3.
Check that the water trap container is empty and properly
attached.
Empty the water trap container whenever the container is more
than half full. Under normal conditions, the D-fend Pro container
fills in 24 hours.
Before connecting the exhaust line to the sample gas
outlet on the airway module, ensure the other end is
connected to the sample gas return port on the
anesthesia machine. Incorrect connections may cause
patient injury.
4.
Connect a line from the airway module sample gas outlet to the
sample gas return port on the anesthesia machine.
5.
Attach the gas sampling line to the sampling line connector on
the water trap.
6.
Turn the system on. The system does a series of automated self
tests. The automatic agent identification is activated.
7.
Connect the sampling line to the airway adapter. Take the gas
sample as close to the patient’s airway as possible. Position the
adapter’s sampling port upwards to prevent condensed water
from entering the sampling line.
1
2
3
AC.24.064
1.
Airway module
2.
Gas sampling line
3.
Airway adapter with sampling line connector
Figure 6-3 • Airway gases setup with airway module
WARNING
Carestation
™
620/650/650c (A1)
6-6
2076152-001
Содержание Carestation 620
Страница 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Страница 12: ...Carestation 620 650 650c A1 x 2076152 001...
Страница 57: ...3 Push the ComWheel or select the quick key to confirm the change 2 System controls and menus 2076152 001 2 23...
Страница 58: ...2 24 2076152 001...
Страница 98: ...3 40 2076152 001...
Страница 102: ...4 4 2076152 001...
Страница 128: ...6 12 2076152 001...
Страница 150: ...7 22 2076152 001...
Страница 178: ...4 Tighten all thumb screws on the rear of the mounting plate AC 24 265 Carestation 620 650 650c A1 8 28 2076152 001...
Страница 182: ...8 32 2076152 001...
Страница 190: ...12 Calibrate the new O2 cell according to the Circuit O2 cell procedure Carestation 620 650 650c A1 9 8 2076152 001...
Страница 196: ...9 14 2076152 001...
Страница 246: ...11 44 2076152 001...
Страница 250: ...3 Set the cost for each agent 4 Select Back to return to the Gas Usage menu Carestation 620 650 650c A1 12 4 2076152 001...
Страница 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...