Individual tests
The
Full Test
or all of the individual tests must be completed with
passing results at least once within every 24-hour period.
Individual tests allow the user to perform any combination of single
tests. These tests are helpful if there is a specific problem/alarm and
the user wishes to test only that portion of the system.
The tests do not automatically move on to the next test. After
completing a test, do another or start a case. If a test fails, follow the
instructions to perform a recheck or accept the results.
The airway module sampling flow is paused during the checkout
when an airway module is installed within the system. If an airway
module is installed on an external monitor outside of the system,
disconnect the gas sampling line from the patient circuit and close
the sample port before performing the checkout.
Ventilator leak
The
Ventilator Leak
test checks the ventilator, bellows, gas
supplies, and breathing circuit for gas leaks in the mechanical
ventilation mode. The patient circuit compliance is also measured in
this test.
To run this test, follow the on-screen instructions.
Vaporizer leak
The
Vaporizer Leak
test is a two-step test. The test first checks the
ventilator, bellows, gas supplies, and then checks for leaks in each
vaporizer.
To run this test, follow the on-screen instructions.
Gas controls
The
Gas Controls
test checks the flow control valves and gas flows.
To run this test, follow the on-screen instructions.
Circuit leak
The
Circuit Leak
test checks the Bag/Vent switch, proper gas supply
pressures, airway pressure measurement transducer, APL valve,
and manual circuit leak.
Note
Carestation
™
620/650/650c (A1)
5-10
2076152-001
Содержание Carestation 620
Страница 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Страница 12: ...Carestation 620 650 650c A1 x 2076152 001...
Страница 57: ...3 Push the ComWheel or select the quick key to confirm the change 2 System controls and menus 2076152 001 2 23...
Страница 58: ...2 24 2076152 001...
Страница 98: ...3 40 2076152 001...
Страница 102: ...4 4 2076152 001...
Страница 128: ...6 12 2076152 001...
Страница 150: ...7 22 2076152 001...
Страница 178: ...4 Tighten all thumb screws on the rear of the mounting plate AC 24 265 Carestation 620 650 650c A1 8 28 2076152 001...
Страница 182: ...8 32 2076152 001...
Страница 190: ...12 Calibrate the new O2 cell according to the Circuit O2 cell procedure Carestation 620 650 650c A1 9 8 2076152 001...
Страница 196: ...9 14 2076152 001...
Страница 246: ...11 44 2076152 001...
Страница 250: ...3 Set the cost for each agent 4 Select Back to return to the Gas Usage menu Carestation 620 650 650c A1 12 4 2076152 001...
Страница 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...