Circuit compliance compensation
The ventilator adjusts gas delivery and monitoring to compensate for
the compliance of the patient circuit if:
•
The
Circuit Compliance
is set to
On
from
Super User
-
System Config.
-
Ventilator Settings
.
•
The
Checkout
is completed after the system is turned on.
Circuit compliance is measured after
Vaporizer Leak
test during the
Checkout
-
Full Test
or when the
Ventilator Leak
check is done as
an individual test.
Circuit Compliance
must be set by the Super
user for the Circuit Compliance compensation function to be active.
See "
" section for information on the circuit
compliance setting.
In volume modes, circuit compliance compensation increases the
volume delivered at the inspiratory port by taking into consideration
the circuit compliance value. In all modes, circuit compliance
compensation adjusts the volume measurements. Circuit compliance
compensation provides consistent ventilator accuracy at the patient
circuit.
Breathing circuits and breathing circuit components are available in
many different configurations from multiple suppliers. Attributes of
the breathing circuits such as materials, tube length, tube diameter,
and configuration of components within the breathing circuit, may
result in hazards to the patient from increased leakage, added
resistance, or changed circuit compliance. It is recommended that a
test be conducted prior to use with each patient.
Perform a
Full Test
or the individual
Ventilator Leak
test
after changing the patient tube type. Changing the patient
breathing circuit after completing a
Full Test
or the
individual
Ventilator Leak
test affects the volume
measurements in all modes.
WARNING
5 Preoperative tests
2076152-001
5-5
Содержание Carestation 620
Страница 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Страница 12: ...Carestation 620 650 650c A1 x 2076152 001...
Страница 57: ...3 Push the ComWheel or select the quick key to confirm the change 2 System controls and menus 2076152 001 2 23...
Страница 58: ...2 24 2076152 001...
Страница 98: ...3 40 2076152 001...
Страница 102: ...4 4 2076152 001...
Страница 128: ...6 12 2076152 001...
Страница 150: ...7 22 2076152 001...
Страница 178: ...4 Tighten all thumb screws on the rear of the mounting plate AC 24 265 Carestation 620 650 650c A1 8 28 2076152 001...
Страница 182: ...8 32 2076152 001...
Страница 190: ...12 Calibrate the new O2 cell according to the Circuit O2 cell procedure Carestation 620 650 650c A1 9 8 2076152 001...
Страница 196: ...9 14 2076152 001...
Страница 246: ...11 44 2076152 001...
Страница 250: ...3 Set the cost for each agent 4 Select Back to return to the Gas Usage menu Carestation 620 650 650c A1 12 4 2076152 001...
Страница 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...