Entropy
WARNING
Ensure that the sensor is placed on the patient’s forehead
according to the instructions. Placing the sensor in a way
other than instructed might result in risks during patient
defibrillation.
Entropy cautions
CAUTION
Strong 30-40 Hz magnetic fields may cause erroneous
Entropy measurement. Do not use devices with such a field
close to the module or sensor.
CAUTION
Diathermy and external interference may degrade the
performance.
CAUTION
The Entropy measurement is always to be used only as an
adjunct to other physiological parameters. Clinicians are
advised to use their knowledge and experience when making
clinical judgements. Entropy values are not to be used as sole
indicators of the patient status.
CAUTION
Check the sensor expiration date on the sensor package. Do
not use expired sensors.
CAUTION
Do not use a sensor for more than 24 hours.
CAUTION
Long-term use of the electrodes may degrade condition of the
skin especially on patients with liver diseases.
CAUTION
Automatic sensor check may need to be disabled if the 70
Hz impedance check signal interferes with other equipment,
such as EEG module with evoked potentials measurement.
CAUTION
Allow the cable to dry completely after cleaning. Moisture and
dirt on the connector can affect the measurement accuracy.
Entropy indications for use
The GE Healthcare Entropy module, E-ENTROPY, and accessories are indicated for
adult and pediatric patients older than 2 years within a hospital for monitoring the
state of the brain by data acquisition of electroencephalograph (EEG) and frontal
electromyograph (FEMG) signals. The spectral entropies, Response Entropy (RE) and
State Entropy (SE), are processed EEG and FEMG variables. The Entropy measurement
is to be used as an adjunct to other physiological parameters.
In adult patients, Response Entropy (RE) and State Entropy (SE) may be used as an aid
in monitoring the effects of certain anesthetic agents, which may help the user titrate
anesthetic drugs according to the individual needs of adult patients. Furthermore
in adults, the use of Entropy parameters may be associated with a reduction of
anesthetic use and faster emergence from anesthesia.
The Entropy module is indicated for use by qualified medical personnel only.
Entropy measurement limitations
●
This measurement is available in OR and PACU software packages only.
350
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...