CO2 with CAPNOSTAT Mainstream, CapnoFlex LF, and Dual CO2 modules
WARNING
With deactivated
Apnea
alarm, keep the patient under close
surveillance.
Apnea alarms’ deactivation with the pause audio key
Apnea alarms can be deactivated with the pause audio key if the
Allow alarm
deactivation with the Audio Pause key for:
setting
Apnea (CO2/Imped)
is enabled in
the
Care Unit Settings
. This setting is password protected.
For more information, see the CARESCAPE Modular Monitors Supplemental
Information Manual.
Selecting CO
2
average
You can select a time interval for averaging the CO
2
.
1.
Select a gas related parameter window.
2.
Select the
CO2
tab >
Setup
.
3.
Select an option from the
CO2 Average
list.
Selecting the FiO
2
level
The presence of a large concentration of oxygen causes the CO
2
level appear lower
than the actual value. Use this option to compensate for the presence of O
2
.
1.
Select a gas related parameter window.
2.
Select the
CO2
tab >
Setup
.
3.
Select an option from the
FiO2 level
list.
Selecting the N
2
O level
NOTE
With the anesthetic agent measurement license only.
Available for OR, PACU or ICU software packages only.
The presence of N
2
O causes the CO
2
value to appear higher than the actual value.
Use this option to compensate for the presence of N
2
O.
1.
Select a gas related parameter window.
2.
Select the
CO2
tab >
Setup
.
3.
Select an option from the
N2O level
list.
Selecting the pump on or off
The pump must be on when using the sidestream method. Note also that a blocked
line may cause the pump to shut off automatically.
Pump on or off with the CapnoFlex LF module
The pump is automatically turned on when a sample line is inserted into the CapnoFlex
LF module. The pump turns off when the sample line is removed.
Selecting pump on or off with the Dual CO2 module
1.
Select a gas related parameter window.
2.
Select the
CO2
tab >
Setup
.
316
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...