Invasive pressures
Displayed values will switch between: Art 123/51 ( ) and Art 100/60 ( )
Invasive pressure troubleshooting
Problem
Solution
Artifact detected and the
Smart BP
option is turned
on.
●
Check the patient.
●
Reposition the catheter.
●
Zero the transducer.
●
If problem persists, turn off the
Smart BP
option.
●
If the
Smart BP
option is turned off, use the audio
pause feature before drawing blood and before
zeroing to reduce nuisance alarms.
Invasive pressure readings seem unstable.
●
Make sure there are no air bubbles in the
transducer systems.
●
Flush and zero.
●
Place the transducer on the patient’s phlebostatic
axis.
---/--- (80) Systolic and diastolic pressure values do
not display.
This may be due to the
Smart BP
option detecting
artifact. When artifacts are detected, only
Mean
values are displayed.
●
Check the patient.
●
Turn off the
Smart BP
if required.
Invasive pressure waveform is displayed but no
numeric values are displayed.
●
Zero the channel. Invasive pressure numeric
values are displayed only for successfully zeroed
channels.
Zeroing of invasive pressure channel(s) fails.
●
Ensure that the channels are open to air.
The channel standby is not selectable, or it is
terminated without the user giving the activation
request.
If the pressures remain between 10 mmHg and
250 mmHg for 10 seconds or more, the selection is
disabled or the standby is terminated.
252
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...