Invasive pressures
5.
To record a realtime PA wedge waveform during analysis, select
Print PA
Waveform
. To stop printing, select
Stop Printing
or
Cancel Printing
.
6.
Inflate the catheter balloon when the
Inflate the Balloon
message displays.
NOTE
PDM and TRAM only: Once the PCWP waveform is
detected, the
Wedge processing
message displays.
After 10 seconds, the automated wedge program displays the
Deflate the
balloon
, followed by the
Wedge Complete
message.
7.
To adjust the PA wedge value, move the cursor up or down with the
PCWP
/ Cursor
arrows.
8.
To save the PCWP value, select
Confirm Wedge
.
The saved PA wedge value displays in the parameter window and is stored in
trends and the hemodynamic calculations.
9.
To print a PCWP report, select
Print PA Report
. To stop printing, select
Waveforms
or
Cancel Printing
.
The PA wedge report contains 20 seconds of waveform data displayed at a
waveform speed of 12.5 mm/s.
Starting a new PA wedge measurement
You can clear the current wedge measurement and start a new one:
1.
Select PA invasive pressure parameter window.
2.
Select
Setup
.
3.
Select
Wedge
.
4.
Select
Restart Wedge
.
Other selections in the
Wedge
menu
There are also two other selections in the
Wedge
menu:
●
C.O.
: This selection will open the C.O.
Setup
menu.
●
Calculations
: This selection will open the
Hemo
calculations menu.
Calibrating the invasive pressure measurement
with PDM and TRAM
1.
Connect all invasive pressure transducers being tested to the module.
2.
Select
Monitor Setup
>
IP Calibration
.
3.
Select the pressure site to test.
4.
Increase or decrease the pressure to adjust the displayed pressure to a known
calibration factor and select
Confirm
.
5.
Repeat for each pressure site as needed.
6.
Select
Previous Menu
>
Close
.
248
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...