Impedance respiration
WARNING
— DEFIBRILLATOR PRECAUTIONS — Patient signal inputs
labeled with the CF and BF symbols with paddles are
protected against damage resulting from defibrillation
voltages. To ensure proper defibrillator protection, use only
the recommended cables and leadwires.
WARNING
— ELECTRODE CONFIGURATION — Impedance respiration
monitoring is not reliable when ECG electrodes are placed
anywhere but on the chest.
WARNING
— ELECTRODES — Whenever patient defibrillation is
a possibility, use non-polarizing (silver/silver chloride
construction) electrodes for ECG monitoring. Polarizing
electrodes (stainless steel or silver constructed) may cause
the electrodes to retain a residual charge after defibrillation. A
residual charge will block acquisition of the ECG signal.
WARNING
— PDM and TRAM — If the
Cardiac Artifact Alarm
is turned
off, apnea events may not be detected.
Respiration cautions
CAUTION
The impedance respiration measurement may cause rate
changes in Minute Ventilation Rate Responsive Pacemakers.
Set the pacemaker rate responsive mode off (PDM and TRAM
only) or turn off the impedance respiration measurement on
the monitor.
Respiration measurement limitations
●
PSM: Impedance respiration is intended for patients over three years old.
●
Electrical devices, such as electrosurgery units and infrared heaters, that emit
electromagnetic disturbance, may cause artifacts or disable the respiration
measurement completely.
●
Movement artifacts, shivering, and interference from the heart may interfere with
the respiration measurement.
Respiration points to note
●
Connect only one impedance respiration module to the monitor simultaneously.
●
Do not place electrodes on obvious layers of fat, or major muscles.
●
Make sure the electrode gel is moist.
●
Make sure electrodes have good skin contact.
●
Depending on the respiration module used, not all respiration measurements and
settings are available to view or change.
●
Since respiration monitoring is so closely linked with ECG monitoring, patient
preparation and electrode placement are important.
●
Intermittent mechanical ventilation: During spontaneous breathing the ventilator
may at times support the patient’s ventilation with an extra inspiration. If these
ventilator inspirations are substantially larger than the spontaneous breaths, the
respiration calculation may mistakenly count only the inspirations and expirations
produced by the ventilator.
192
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...