ECG
When the
Maximum
filter is selected, the prompt text
Warning: Maximum
filter alters the displayed ECG morphology
is displayed.
Setting the QRS width
NOTE
This setting affects the arrhythmia detection sensitivity.
This selection is not available with combination monitoring.
If the
QRS Width
is locked in the
Care Unit Settings
, the option is not selectable.
1.
Select the HR parameter window.
2.
Select the
Advanced
tab.
3.
Select a setting from the
QRS Width
list. Choices are:
●
Narrow
: Intended for use with all neonates and the pediatric patient with a
QRS complex width of 100 ms or less. This is the default setting for the
Infant
and
Pediatric
profiles.
●
Normal
: Intended for ECG rhythms that have QRS complex widths of
approximately 70 ms or wider (for example, almost all adult patients and any
patient with electronic ventricular pacing).
Selecting the leads for ECG analysis
You can choose whether the monitor performs an ECG analysis using single lead ECG
data or data from multiple ECG leads. Multiple ECG leads will typically reduce false
alarms and improve the detection sensitivity. However, if most leads are noisy or low
amplitude, the
Single lead
mode using the best available ECG lead will help.
With a 3-leadwire cable the setting is
Single lead
and cannot be changed. If the
measurement mode is changed from the 3–leadwire mode to 5–, 6–, 10–lead or 12RL
mode, the setting changes to
Multi lead
.
1.
Select the HR parameter window.
2.
Select the
Advanced
tab.
3.
Select an option from the
Lead Analysis
list. The choices are:
●
Single lead
: EkPro algorithm uses one of the leads I, II, III, or V1 for the analysis.
ECG 1 Lead
is used for the analysis if it is I, II, III, or V1. If it is anything else, then
the following mapping is used: V2 to V6 = V1, aVR = II, aVL = I, aVF = III. Also
note that the ST values are only calculated for the single lead.
●
Multi lead
: EkPro algorithm uses the following leads:
■
3–lead mode: the only measured lead (I, II, or III)
■
5–lead and 6–lead mode: any lead assigned to Va.
■
12RL mode: I, II, III, and V1.
■
12–lead mode: I, II, III, and V1.
Relearning the patient’s QRS pattern
During ECG monitoring, you may need to use the
Relearn QRS
feature when a
dramatic change in the patient’s ECG pattern has occurred. Allowing the monitor to
learn the new ECG pattern corrects false arrhythmia alarms and heart rate values,
and restores the ST measurements. Relearning takes typically 30 seconds or less. The
message
Relearning...
displays while the monitor relearns the QRS pattern. During this
2062971-001
CARESCAPE Modular Monitors
161
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...