ECG
NOTE
The
ECG Size
setting affects arrhythmia detection and heart
rate calculation sensitivity. Normal waveform size/QRS
detection sensitivity is
1x
. Size
2x
and greater increases
the QRS detection sensitivity. This may be helpful for low
amplitude QRS waveforms. Use with caution since baseline
artifact may be detected as a QRS complex.
Selecting the hemodynamic waveform sweep speed
NOTE
This setting adjusts the waveform speed for all of the
hemodynamic parameters.
1.
Select the HR parameter window.
2.
Select a numeric value from the
Hemodynamic Sweep Speed
list.
The smaller the value, the slower the sweep speed.
Printing all ECG waveforms
NOTE
Pressing Graph in the telemetry transmitter will start the
telemetry waveform printing.
1.
Select the HR parameter window.
2.
Select
All ECG Waveforms
.
3.
Select
Print Page
.
4.
You can stop printing by selecting
Stop Printing
or
Cancel Printing
.
Selecting the ECG waveform filter
You can select how the waveform appears on the display and on the printout.
1.
Select the HR parameter window.
2.
Select the
Advanced
tab.
3.
Select a filter from the
Waveform Filter
list. Choices are:
●
Diagnostic
:
■
TRAM: 0.05 to 100 Hz.
♦
TRAM with a 10-leadwire cable: The waveform filter is automatically set
to
Diagnostic
and cannot be changed.
■
E-modules and PDM: 0.05 Hz to 150 Hz.
●
Monitoring
:
■
PDM, E-modules, TRAM: 0.05 Hz to 32 Hz (with 50 Hz powerline frequency).
■
PDM, E-modules, TRAM: 0.05 Hz to 40 Hz (with 60 Hz powerline frequency).
■
Telemetry transmitters: 0.05 Hz to 40 Hz. The waveform filter is
automatically set to
Monitoring
and cannot be changed.
●
Moderate
:
■
PDM, E-modules, TRAM: 0.05 Hz to 25 Hz.
●
Maximum
:
■
PDM, E-modules, TRAM: 5 Hz to 25 Hz.
160
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...