ECG
Setting the beep tone during bradycardia and HR low
alarms
NOTE
PDM and TRAM only. NICU software package only.
This selection is not available with combination monitoring.
This selection is available if the
Beat Volume
for QRS is set to 0 (off).
If the alarm for bradycardia has been set to off or silenced, or the ECG alarms are
silenced permanently, then the QRS tone does not sound, either.
1.
Select the HR parameter window.
2.
Set the beep tone:
Beat Tone on Brady Only
>
On
or
Off
.
When the beep tone is selected
On
, the QRS tones will sound only with
Brady
alarm conditions.
●
If the alarm volume has been set to below 8, the QRS tones will sound at the
selected alarm volume level +2.
●
If the alarm volume has been set to 8 or more, the QRS tones will sound at
alarm volume level 10.
Variable beat tone
You can configure a variable beat tone through
Monitor Setup
>
Default Setup
>
Care
Unit Settings
>
Parameters
>
Variable Beat Tone
. This setting is password protected.
If it set to
All beat sources
, the SpO
2
saturation affects all beep sounds including ECG
and IP when the SpO
2
measurement is available: beep frequency changes according
to increasing and decreasing SpO
2
values. If the setting is set to
Only SpO2
, other beep
sounds are not affected by the changing SpO
2
values.
For more information, see the CARESCAPE Modular Monitors Supplemental
Information Manual.
Aspect ratio and different display sizes
ECG aspect ratio is the ratio of the vertical sensitivity (ECG size) to the horizontal
sensitivity (sweep speed).
The B450 aspect ratio is optimized for the 12” integrated display, and the B650 aspect
ratio is optimized for the 15” integrated display. If larger external displays are used,
the waveform size and sweep speed may change.
The B850 aspect ratio is optimized for a 19” display. If a display of a different size is
used, the waveform size and sweep speed may change.
For more information, see the CARESCAPE Modular Monitors Supplemental
Information Manual.
Selecting the ECG waveform size
This selection adjusts the size of the displayed ECG waveform.
1.
Select the HR parameter window.
2.
Select a value from the
ECG Size
list.
The selections are
0.5x
,
1x
,
2x
,
4x
. The smaller the value, the smaller the waveform.
2062971-001
CARESCAPE Modular Monitors
159
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...