Alarms
3.
Select
Default Limits
.
If you need to undo these changes, select
Undo Settings
before closing the menu.
Turning off all local alarm indicators (sleep mode)
NOTE
This feature is not available with OR and PACU software
packages.
If the feature has been enabled in the
Care Unit Settings
(password protected), you
can turn off the monitor’s display and turn off all audible, visual, and alarm light alarm
indicators until you turn them back on again. Patient monitoring is occurring; however,
the monitor is not displaying patient data or indicating patient alarms locally. Local
printing is also inactive. Alarms are logged and trended. If the monitor is connected to
the network, alarms and alarm printouts, and parameter data (and nurse call signals
with B850) will continue to be sent over the network during sleep mode. For more
information on enabling or disabling the
Audio&Display
>
Off Allowed
feature, see the
CARESCAPE Modular Monitors Supplemental Information Manual.
1.
Select
Alarm Setup
from the monitor’s main menu.
2.
Select the
Audible & Visual
tab.
3.
Select
Audio&Display
>
Off
.
A screen saver replaces the display of patient data. Any user input (touchscreen,
button, Trim Knob, keyboard key, mouse click) reactivates the alarms and the
monitoring screen.
Remote management of alarms
The remote alarm settings are defined in the
Care Unit Settings
and they are
password protected. The following settings are available:
●
Allowing remote audio pause from this monitor for a remote bed.
●
Allowing audio pause for this monitor from a central station or remote monitor. Not
available with the OR software package.
●
Allowing remote pausing of different alarm priorities.
●
Showing the remote patient name.
●
Turning the remote monitor’s alarm light on or off.
●
Selecting the remote alarm notification tone.
For more information, see the CARESCAPE Modular Monitors Supplemental
Information Manual.
Alarm settings after a power loss
If the monitor loses power, the amount of time without power affects whether or
not you need to reset the alarm settings.
144
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...